CLASS II Device recall · Reported 9 Sep 2015 · FDA Enforcement Report

KimVent* Closed Suction System for Adults, 14 F, DSE, MDI (Product recalled

Halyard Health is recalling KimVent* Closed Suction System for Adults, 14 F, DSE, MDI (, distributed nationwide in the US. The recall was initiated on 29 Jul 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Code 22108 - M5082T626, M5089T625; Product Code 221038 - M5089T606, M5089T631
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2508-2015
FDA event ID
71966
Recalling firm
Halyard Health, Alpharetta, GA
Classification
Class II
Status
Terminated
Recall initiated
29 Jul 2015
FDA classified
31 Aug 2015
Reported
9 Sep 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

29 Jul 2015Recall initiated by company
31 Aug 2015Classified by FDA+33 days
9 Sep 2015Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Product as listed by the FDA

KimVent* Closed Suction System for Adults, 14 F, DSE, MDI (Product Code 22108); KimVent* Closed Suction System for Adults, 14 F, DSE, MDI (Product Code 221038) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Where it was distributed

Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Which lots are affected?

Product Code 22108 - M5082T626, M5089T625; Product Code 221038 - M5089T606, M5089T631

Where was it sold?

Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

31 other products · FDA event 71966

This product is part of a recall event; the main page for the event is NovaPlus Trach Care* Closed Suction System for Adults recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IINovaPlus Trach Care* Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Turbo-Cleaning Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, DSE (Product recalledHalyard HealthOtherNationwide
CLASS IINovaPlus Trach Care* Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 12 F, Elbow (Product recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Wet Pak* Closed Suction System for Pediatrics recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Neonates/Pediatrics, 6 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Wet Pak* Closed Suction System for Adults, 14 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Turbo-Cleaning Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, T-Piece recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Wet Pak* Closed Suction System for Adults, 14 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, D-Tip recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 12 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Turbo-Cleaning Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Turbo-Cleaning Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Multi-Access Port Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Wet Pak* Closed Suction System for Adults, 14 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Multi-Access Port Replacement Catheter recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, DSE, DT (Product recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction Systems for Adults, 14 F, Qwik Clip (Product recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, MDI recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 12 F, T-Piece (Product recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Multi-Access Port Replacement Catheter for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Multi-Access Port Closed Suction System recalledHalyard HealthOtherNationwide
CLASS IINovaPlus Trach Care* Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IINovaPlus Trach Care* Closed Suction System for Adults recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Neonates/Pediatrics, 5 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Neonates/Pediatrics, 5 F recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Closed Suction System for Adults, 14 F, Elbow recalledHalyard HealthOtherNationwide
CLASS IIKimVent* Wet Pak* Closed Suction System for Adults, 14 F recalledHalyard HealthOtherNationwide
CLASS IINovaPlus Trach Care* Closed Suction System for Adults recalledHalyard HealthOtherNationwide

More recalls from Halyard Health

All 57

Other Heart & Vascular Devices recalls

All 6,199