CLASS II ONGOING Device recall · Reported 18 Feb 2026 · FDA Enforcement Report

Laerdal Compact Suction Unit 4 RTCA recalled by Laerdal Medical

Laerdal Medical is recalling Laerdal Compact Suction Unit 4 RTCA, distributed nationwide in the US. The recall was initiated on 13 Jan 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;
Quantity recalled1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1341-2026
FDA event ID
98039
Recalling firm
Laerdal Medical, Wappingers Falls, NY
Classification
Class II
Status
Ongoing
Recall initiated
13 Jan 2026
FDA classified
11 Feb 2026
Reported
18 Feb 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

13 Jan 2026Recall initiated by company
11 Feb 2026Classified by FDA+29 days
18 Feb 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Product as listed by the FDA

Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.

Nationwide

Questions about this recall

Why was it recalled?

units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Which lots are affected?

1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Laerdal Medical

All 3
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IILcsu 4, Laerdal Compact Suction Unit The Lcsu recalledLaerdal MedicalOtherNationwide
DeviceCLASS IILaerdal Compact Suction Unit ¿ 4 under the following recalledLaerdal MedicalOtherNationwide

Other Other Medical Devices recalls

All 13,908