CLASS II ONGOING Device recall · Reported 29 May 2019 · FDA Enforcement Report

Leksell GammaPlan recalled by Elekta Instrument AB

Elekta Instrument AB is recalling Leksell GammaPlan 11.1, distributed nationwide in the US. The recall was initiated on 27 Jun 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Version 11.1
Quantity recalled52 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1412-2019
FDA event ID
82662
Recalling firm
Elekta Instrument AB, Stockholm, N/A
Classification
Class II
Status
Ongoing
Recall initiated
27 Jun 2018
FDA classified
23 May 2019
Reported
29 May 2019
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

27 Jun 2018Recall initiated by company
23 May 2019Classified by FDA+330 days
29 May 2019Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.

Product as listed by the FDA

Leksell GammaPlan 11.1 Product Usage: Leksell GammaPlan is a computer-based system designed for Leksell Gamma Knife treatment planning.

Where it was distributed

Worldwide Distribution - US Nationwide distribution to GA, MD, NJ, PA, PR, WA. International distribution to Australia, Belgium, France, Germany, Hong Kong, Italy, Japan, Korea, Democratic People's Republic of, Netherlands, Romania, Russian Federation, Spain, Turkey, United Kingdom.

Nationwide GeorgiaMarylandNew JerseyPennsylvaniaPuerto RicoWashington

Questions about this recall

Why was it recalled?

The margin tool in Leksell GammaPlan 11.1 systematically overestimates margin in certain areas of the volume.

Which lots are affected?

Software Version 11.1

Where was it sold?

Worldwide Distribution - US Nationwide distribution to GA, MD, NJ, PA, PR, WA. International distribution to Australia, Belgium, France, Germany, Hong Kong, Italy, Japan, Korea, Democratic People's Republic of, Netherlands, Romania, Russian Federation, Spain, Turkey, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Elekta Instrument AB

All 5

Other Other Medical Devices recalls

All 13,908