CLASS II Device recall · Reported 30 Nov 2016 · FDA Enforcement Report
Liaison Estradiol II Gen recalled by Diasorin
Diasorin is recalling Liaison Estradiol II Gen, 310680, 310680-CN The DiaSorin Liaison Estradiol II Gen assay, distributed nationwide in the US. The recall was initiated on 13 Jul 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0645-2017
- FDA event ID
- 75478
- Recalling firm
- Diasorin, Stillwater, MN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 13 Jul 2016
- FDA classified
- 21 Nov 2016
- Reported
- 30 Nov 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Potency & Assay
- Device type
- Test Kits & Diagnostics
Timeline
| 13 Jul 2016 | Recall initiated by company | |
| 21 Nov 2016 | Classified by FDA | +131 days |
| 30 Nov 2016 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
DiaSorin is issuing an urgent medical device correction for the LIAISON¿ Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON¿ Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex).
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
LIAISON Estradiol II Gen, 310680, 310680-CN The DiaSorin LIAISON Estradiol II Gen assay is a chemiluminescent immunoassay (CLIA) intended for the quantitative determination of estradiol in human serum. Assay results should be used in conjunction with other clinical and laboratory data to assist in the diagnosis and management of conditions involving an excess or deficiency of estradiol. The test has to be performed on the LIAISON Analyzer Family.
Where it was distributed
USA (nationwide) Distribution and Worldwide Distribution including the states of : AR, AZ, CA, FL, ID, NJ, NV, NY and PA.., and to the countries of : Saudi Arabia, Colombia, Mauritius, Iraq, Mongolia, Greece, Albania, Austria, Bosnia, Bosnia Herzegovina, Belarus, Bulgaria, Macedonia, Poland, Serbia, Dominican Rep., Tunisia, Paraguay, Hungary, Jordan, Egypt, Belgium, Caribbean, Germany, Brazil, Mexico, Italy, India, South Korea, Cyprus, Bangladesh, Morocco, Chile, Bahrain, Palestine, Utd.Arab.Emir., Kosovo, Uruguay, Venezuela, Hungary, Australia, Vietnam, Sri Lanka, Turkey, Vietnam, Romania, Mauritius, Ecuador, Lebanon, Nepal, Argentina, Hong Kong, China and Canada.
Nationwide ArkansasArizonaCaliforniaFloridaIdahoNew JerseyNevadaNew YorkPennsylvania
Questions about this recall
Is Liaison Estradiol II Gen, 310680, 310680-CN The DiaSorin Liaison Estradiol II Gen assay recalled?
Yes. Diasorin recalled Liaison Estradiol II Gen, 310680, 310680-CN The DiaSorin Liaison Estradiol II Gen assay on 13 Jul 2016. The recall is Class II and its status is Terminated. Recall number Z-0645-2017.
Why was it recalled?
DiaSorin is issuing an urgent medical device correction for the LIAISON¿ Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON¿ Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex).
Which lots are affected?
Catalog Number 31 0680, 31 0680-CN
Where was it sold?
USA (nationwide) Distribution and Worldwide Distribution including the states of : AR, AZ, CA, FL, ID, NJ, NV, NY and PA.., and to the countries of : Saudi Arabia, Colombia, Mauritius, Iraq, Mongolia, Greece, Albania, Austria, Bosnia, Bosnia Herzegovina, Belarus, Bulgaria, Macedonia, Poland, Serbia, Dominican Rep., Tunisia, Paraguay, Hungary, Jordan, Egypt, Belgium, Caribbean, Germany, Brazil, Mexico, Italy, India, South Korea, Cyprus, Bangladesh, Morocco, Chile, Bahrain, Palestine, Utd.Arab.Emir., Kosovo, Uruguay, Venezuela, Hungary, Australia, Vietnam, Sri Lanka, Turkey, Vietnam, Romania, Mauritius, Ecuador, Lebanon, Nepal, Argentina, Hong Kong, China and Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Diasorin
All 6| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Liaison Q.s.e.t recalled by Diasorin | Diasorin | Packaging Defects | MN |
| Device | CLASS II | Liaison Hsv-2 Type Specific IgG recalled by Diasorin | Diasorin | Potency & Assay | Nationwide |
| Device | CLASS II | Liaison 1-84 PTH Assay recalled by Diasorin | Diasorin | Potency & Assay | Nationwide |
| Device | CLASS II | Parvovirus B19 IgG EIA kit recalled by Diasorin | Diasorin | Other | 11 states |
| Device | CLASS II | Pro-Trac II Tacrolimus Elisa kit recalled by Diasorin | Diasorin | Potency & Assay | CA |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |