CLASS II Device recall · Reported 8 Apr 2020 · FDA Enforcement Report
LifePAK 500 Automated External Defibrillator recalled by Physio-Control
Physio-Control is recalling LifePAK 500 Automated External Defibrillator, distributed nationwide in the US. The recall was initiated on 20 Mar 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1633-2020
- FDA event ID
- 85194
- Recalling firm
- Physio-Control, Redmond, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 20 Mar 2020
- FDA classified
- 1 Apr 2020
- Reported
- 8 Apr 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Pacemakers & Defibrillators
Timeline
| 20 Mar 2020 | Recall initiated by company | |
| 1 Apr 2020 | Classified by FDA | +12 days |
| 8 Apr 2020 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in a delay in treatment
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
LifePAK 500 Automated External Defibrillator, model numbers: 3005400-XXX, U3005400-XXX, 3011790-XXX/, U3011790-XXX - Product Usage: The LIFEPAK 500 AED is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient s electrocardiographic (ECG) rhythm and indicates whether it detects a shockable rhythm. The LIFEPAK 500 AED requires operator interaction to defibrillate the patient.
Where it was distributed
Worldwide distribution - US Nationwide distribution and countries of Afghanistan, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Barbados, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Cayman Islands, Chile, China, Columbia, Congo, Cook Islands, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Macedonia, the former Yugoslav Republic of, Madagascar, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Province of China, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam, Virgin Islands, British, Virgin Islands, U.S., Zimbabwe.
Questions about this recall
Why was it recalled?
Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient is connected which could result in a delay in treatment
Which lots are affected?
All Lots
Where was it sold?
Worldwide distribution - US Nationwide distribution and countries of Afghanistan, American Samoa, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Barbados, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Cayman Islands, Chile, China, Columbia, Congo, Cook Islands, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Korea, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lithuania, Macedonia, the former Yugoslav Republic of, Madagascar, Malaysia, Mexico, Netherlands, New Caledonia, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, San Marino, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Province of China, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Bolivarian Republic of, Vietnam, Virgin Islands, British, Virgin Islands, U.S., Zimbabwe.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Physio-Control
All 36| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Lifepak 15, REF: 99577-000025, 99577 recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Physio Control Infant/Child Reduced Energy Defibrillator Electrodes recalled | Physio-Control | Other | Nationwide |
| Device | CLASS II | Lifepak 1000 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Lifepak 15 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Lifepak CR2 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
Other Pacemakers & Defibrillators recalls
All 494| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DermaV Laser System recalled by Lutronic | Lutronic | Other | Nationwide |
| Device | CLASS II | Convenience kits containing select SKUs of Polycarbonate recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits: Loop Pack DYNJ61460C Open Heart recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Acl Pack DYNJ64161C Arthroscopy recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Bib AV Fistula Pack DYNJ58304B recalled | Medline Industries | Device Malfunction | Nationwide |