CLASS II ONGOING Device recall · Reported 13 May 2026 · FDA Enforcement Report
Lifepak 15, REF: 99577-000025, 99577 recalled by Physio-Control
Physio-Control is recalling Lifepak 15, REF: 99577-000025, 99577-000047, 99577-001256, distributed nationwide in the US. The recall was initiated on 13 Apr 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2065-2026
- FDA event ID
- 98418
- Recalling firm
- Physio-Control, Redmond, WA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Apr 2026
- FDA classified
- 4 May 2026
- Reported
- 13 May 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Processing Deviation
- Device type
- Other Medical Devices
Timeline
| 13 Apr 2026 | Recall initiated by company | |
| 4 May 2026 | Classified by FDA | +21 days |
| 13 May 2026 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122
Where it was distributed
US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
Nationwide ColoradoFloridaIndianaKentuckyLouisianaMassachusettsMichiganMinnesotaNorth CarolinaNew JerseyNew YorkOhioOklahomaPennsylvaniaSouth CarolinaTexasVirginiaWashingtonWisconsinWest Virginia
Questions about this recall
Is Lifepak 15, REF: 99577-000025, 99577-000047, 99577-001256 recalled?
Yes. Physio-Control recalled Lifepak 15, REF: 99577-000025, 99577-000047, 99577-001256 on 13 Apr 2026. The recall is Class II and its status is Ongoing. Recall number Z-2065-2026.
Why was it recalled?
Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
Which lots are affected?
Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 0088387387171320071648887706, 0088387387168320081748921846, 0088387387169020092448981398, 0088387387169020102649032288, 0088387387169020102649050140, 0088387387162120110249055931, 0088387387168320122949156402, 0088387387169021011149162067, 0088387387169021011149167463, 0088387387169021013049212578, 0088387387169021020949225869, 0088387387169021020949226060, 0088387387176821021949240566, 0088387387168321033149345530, 0088387387169021042149378981, 0088387387169021042649381781, 0088387387169021042749381576, 0088387387168321081849559854, 0088387387168321081949560127, 0088387387164521091649607224, 0088387387171321102149662806, 0088387377697122070750045009, 0088387387169022090150136719, 0088387387171322090850152289, 0088387387708122091350158293, 0088387387169022101950223934, 0088387387708122112150288760, 0088387387171322120150307353, 0088387387169023062850629377, 0088387387169023072450653772, 0088387387169023072450667197, 0088387387708123072850669746, 0088387387708123082850714419, 0088387387171323082950720588, 0088387387169023112150850129, 0088387387708123122050883524, 0088387388434824052251092313, 0088387387170624121151422867, 0088387387170624122351440213, 0088387387170624122351444039, 0088387388434824052251092313, 0088387387169019100848453060, 0088387387201715090843839264, 0088387387171318121847850793/ 38523489, 39736196, 40465345, 40717931, 40765749, 41130420, 41139184, 41420681, 41446039, 41446693, 41704583, 42221131, 42920956, 42921954, 43166573, 43394891, 43524360, 43914574, 43915122, 44187447, 44252611, 44262721, 44277082, 44282280, 44574340, 44896329, 44962804, 45447229, 45743786, 45834209, 45998142, 46103043, 46303288, 46303092, 46587130, 46602003, 46608783, 46614788, 46813507, 46966267, 47337396, 47401928, 47611441, 47784931, 47786317, 47847661, 47848950, 47845317, 47855371, 47857172, 47858660, 47872868, 48018370, 48247559, 48298704, 48453060, 48455890, 48586351, 48586599, 48651507, 48701689, 48784975, 48887706, 48921846, 48981398, 49032288, 49050140, 49055931, 49156402, 49162067, 49167463, 49212578, 49225869, 49226060, 49240566, 49345530, 49378981, 49381781, 49381576, 49559854, 49560127, 49607224, 49662806, 50045009, 50136719, 50152289, 50158293, 50223934, 50288760, 50307353, 50629377, 50653772, 50667197, 50669746, 50714419, 50720588, 50850129, 50883524, 51092313, 51422867, 51440213, 51444039, 51092313, 48453060, 43839264, 47850793, 38369584. LIFEPAK 1000/0088507469887815020943235807, 0088507469895315031343346860, 0088507469887815111944068667, 0088507469887815111944068669, 0088507469895316011844239775, 0088507469887816041444539726, 0088507469887817061745966118, 0088507469887817061745966121, 0088507469887817061745966122, 0088507469887817061745966126, 0088507469887817061945966131, 0088507469887817062945994506, 0088507469887817062946004311, 0088507469887817063046012479, 0088507469887818060547291148, 0088387381546519080648366102, 0088387381545820092948987376, 0088387381545821011249174469, 0088387381545821081349544565, 0088387381545821090249586700, 0088387381546521122249752235, 0088387381545822012649794992, 0088387381545822032949882696, 0088387381545822032949882698, 0088387381545822032949882819/ 34933095, 35096556, 35184465, 36153943, 36199099, 36560529, 36560531, 37119097, 37226402, 37330295, 38074718, 38074722, 38074727, 38074737, 38074741, 38074744, 38074745, 38074747, 38074748, 38074749, 38077938, 38077939, 38077940, 38077941, 38077944, 38378802, 38513659, 38656117, 38656121, 38656122, 38957375, 38957392, 38957397, 38957399, 38957402, 38972629, 38972677, 39775405, 39775406, 39775407, 39775415, 39964908, 40323073, 40412729, 40541187, 40584085, 40594714, 40707896, 40725568, 40725569, 40725571, 40725572, 40725575, 40725584, 40725582, 40725583, 41306695, 41692754, 41863786, 42504017, 42520778, 42803174, 42921847, 43235807, 43346860, 44068667, 44068669, 44239775, 44539726, 45966118, 45966121, 45966122, 45966126, 45966131, 45994506, 46004311, 46012479, 47291148, 48366102, 48987376, 49174469, 49544565, 49586700, 49752235, 49794992, 49882696, 49882698, 49882819. LIFEPAK 35/0088387377265224042251041190, 0088387377265224041851042010, 0088387377265224042251044581, 0088387376560924071151156570, 0088387376560924071151157072, 0088387376560924071151157449, 0088387376560924080651182632, 0088387376560924091251263526, 0088387376560924121051417585, 0088387376560925020451496956/ 51041190, 51042010, 51044581, 51156570, 51157072, 51157449, 51182632, 51263526, 51417585, 51496956. LIFEPAK 20e/0088387385386320082448935344, 0088387385386320082448935408, 0088387385386321080249537075/ 48935344, 48935408, 49537075
Where was it sold?
US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Physio-Control
All 36| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Physio Control Infant/Child Reduced Energy Defibrillator Electrodes recalled | Physio-Control | Other | Nationwide |
| Device | CLASS II | Lifepak 1000 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Lifepak 15 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Lifepak CR2 Part recalled by Physio-Control | Physio-Control | Processing Deviation | Nationwide |
| Device | CLASS II | Lifepak 20e recalled by Physio-Control | Physio-Control | Other | WV |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |