CLASS II Device recall · Reported 31 Aug 2016 · FDA Enforcement Report
Liver Access and Biopsy Needle Set Biopsy needle kit recalled by Cook
Cook is recalling Liver Access and Biopsy Needle Set Biopsy needle kit Intended for use in obtaining liver, distributed nationwide in the US. The recall was initiated on 15 Apr 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2618-2016
- FDA event ID
- 74010
- Recalling firm
- Cook, Bloomington, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 15 Apr 2016
- FDA classified
- 22 Aug 2016
- Reported
- 31 Aug 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Syringes & Needles
Timeline
| 15 Apr 2016 | Recall initiated by company | |
| 22 Aug 2016 | Classified by FDA | +129 days |
| 31 Aug 2016 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Liver Access and Biopsy Needle Set Biopsy needle kit Intended for use in obtaining liver histology samples via a jugular vein approach.
Where it was distributed
Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
Questions about this recall
Is Liver Access and Biopsy Needle Set Biopsy needle kit Intended for use in obtaining liver recalled?
Yes. Cook recalled Liver Access and Biopsy Needle Set Biopsy needle kit Intended for use in obtaining liver on 15 Apr 2016. The recall is Class II and its status is Terminated. Recall number Z-2618-2016.
Why was it recalled?
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Which lots are affected?
All Lots. Catalog prefixes LABS Catalog/product number; Global/order number LABS-100 G08283 LABS-200 G08284 LABS-200-CHP-A G10590
Where was it sold?
Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cook
All 264| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Skinny Needle with Chiba Tip recalled by Cook | Cook | Other | Nationwide |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Transseptal Needle, Trocar recalled by Cook | Cook | Device Malfunction | Nationwide |
| Device | CLASS II | Transseptal Needle with Catheter recalled by Cook | Cook | Foreign Material | Nationwide |
Other Syringes & Needles recalls
All 1,060| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | BD Needle Kit for Powered Driver recalled by Bard Access Systems | Bard Access Systems | Other | 32 states |
| Device | CLASS II | Boston Scientific Enhance Transcarotid/Peripheral Access Kit recalled | Galt Medical | Other | Nationwide |
| Device | CLASS II | Geneva Brachytherapy Applicator recalled by Elekta | Elekta | Other | Nationwide |
| Device | CLASS II | Brachytherapy Applicator Advanced Gynecological Applicator recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Medline 111 Gravity Set W/3 Needle Free Y-Sites recalled | Bd Switzerland Sarl | Other | IL |