CLASS II Device recall · Reported 12 Jul 2017 · FDA Enforcement Report

Logic Fit Tibial Tamp Head recalled by Exactech

Exactech is recalling Logic Fit Tibial Tamp Head, distributed in 32 states. The recall was initiated on 25 May 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog #: 213-75-01 Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008, 51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and 61872-002.
Quantity recalled573

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2662-2017
FDA event ID
77001
Recalling firm
Exactech, Gainesville, FL
Classification
Class II
Status
Terminated
Recall initiated
25 May 2017
FDA classified
3 Jul 2017
Reported
12 Jul 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 May 2017Recall initiated by company
3 Jul 2017Classified by FDA+39 days
12 Jul 2017Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Logic Fit Tibial Tamp Head

Where it was distributed

AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom

AlabamaArizonaCaliforniaColoradoFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasLouisianaMassachusettsMarylandMaineMinnesotaNorth CarolinaNew JerseyNevada

Questions about this recall

Is Logic Fit Tibial Tamp Head recalled?

Yes. Exactech recalled Logic Fit Tibial Tamp Head on 25 May 2017. The recall is Class II and its status is Terminated. Recall number Z-2662-2017.

Why was it recalled?

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

Which lots are affected?

Catalog #: 213-75-01 Lot #'s: 37166-002, 44944-004, 44944-005, 44944-006, 48073-010, 48073-011, 48073-012, 49036-017, 51180-004, 51180-005, 51180-006, 51180-008, 51754-031, 51754-032, 51754-033, 51754-034, 59743-001, 61872-001,and 61872-002.

Where was it sold?

AL, AZ, CA, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, France, Germany, India, Italy, Japan, Korea, Luxembourg, Singapore, Spain, Switzerland, The Netherlands and United Kingdom

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Exactech

All 197

Other Hip & Knee Implants recalls

All 3,097