CLASS II ONGOING Device recall · Reported 13 May 2026 · FDA Enforcement Report

Lux-dx II, Arrythmia detector and alarm recalled by Boston Scientific

Boston Scientific is recalling Lux-dx II, Arrythmia detector and alarm, distributed nationwide in the US. The recall was initiated on 30 Mar 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787
Quantity recalled2 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2048-2026
FDA event ID
98679
Recalling firm
Boston Scientific, Saint Paul, MN
Classification
Class II
Status
Ongoing
Recall initiated
30 Mar 2026
FDA classified
1 May 2026
Reported
13 May 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Mar 2026Recall initiated by company
1 May 2026Classified by FDA+32 days
13 May 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5

Where it was distributed

Worldwide distribution - US Nationwide and the country of Belgium.

Nationwide

Questions about this recall

Why was it recalled?

For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.

Which lots are affected?

LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787

Where was it sold?

Worldwide distribution - US Nationwide and the country of Belgium.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 98679
ReportedClassRecallCompanyReasonWhere
CLASS IILUX-Dx II Plus, Arrythmia detector and alarm recalledBoston ScientificDevice MalfunctionNationwide

More recalls from Boston Scientific

All 628

Other Other Medical Devices recalls

All 13,908