CLASS II Device recall · Reported 25 Nov 2020 · FDA Enforcement Report

Mako Integrated Cutting System (MICS) Handpiece recalled

Mako Surgical is recalling Mako Integrated Cutting System (MICS) Handpiece, distributed nationwide in the US. The recall was initiated on 23 Oct 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes(01) 00848486030193 (21)<serial number>, Serial numbers: 4200052, 4200095, 4200112, 4200117, 4200119, 4200167, 4200190, 4200193, 4200202, 4200203, 4200204, 4200226, 4200233, 4200235, 4200248, 4200250, 4200257, 4200269, 4200270, 4200288, 4200293, 4200305, 4200350, 4200373, 4200384, 4200387, 4200391, 4200397, 4200398, 4200400, 4200409, 4200436, 4200457, 4200459, 4200477, 4200488, 4200512, 4200540, 4200548, 4200550, 4200558, 4200624, 4200645, 4200653, 4200666, 4200717, 4200766, 4200774, 4200785, 4200789, 4200801, 4200806, 4200807, 4200816, 4200843, 4200856, 4200868, 4200904, 4200917, 4201000, 4201040, 4201068, 4201089, 4201102, 4201103, 4201121, 4201167, 4201218, 4201236, 4201238, 4201253, 4201296, 4201302, 4201366, 4201367, 4201376, 4201385, 4201408, 4201412, 4201442, 4201452, 4201454, 4201477, 4201490, 4201630, 4201661, 4201714, 4201752, 4201829, 4201831, 4201851, 4202009, 4202020, 4202038, 4202044, 4202062, 4202063, 4202122, 4202146, 4202162, 4202175, 4202315, 4202333, 4202626, 4202631, 4202659, 4202729, 4202742, 4202882, 4202987, 4203180, 4203198, 4203289, 4203424, 4203441, 4203649, 4203709, 4203728, 4203730, 4203731, 4203736, 4203807, 4203826, 4204015, 4204164, 4204389, 4204448, 4204539, 4204553, 4204577, 4204673, 4204812, 4204934, 4205093, 4205190, 4205380, 4205426, 4206219, 4206271, 4209023, 4209025, 4209044, 4209045, 4209046, 4209065, 4209066, 4209067, 4209068, 4209071, 4209149, 4209151, 4209155, 4209156, 4209157, 4209162, 4209164, 4209166, 4209167, 4209171, 4209185, 4209189, 4209210, 4209211, 4209213, 4209220, 4209244, 4209250, 4209252, 4209255, 4209256, 4209259, 4209261, 4209262, 4209267, 4209268, 4209271, 4209298, 4209299, 4209303, 4209304, 4209306, 4209308, 4209312, 4209317, 4209318, 4209345, 4209346, 4209357, 4209360, 4209363, 4209364, 4209365, 4209387, 4209388, 4209389, 4209398, 4209441, 4209443, 4209460, 4209463, 4209464, 4209466, 4209471, 4209472, 4209473, 4209474, 4209475, 4209476, 4209477, 4209486, 4209487, 4209491, 4209493, 4209494, 4209495, 4209496, 4209500, 4209519, 4209522, 4209526, 4209529, 4209531, 4209532, 4209533, 4209535, 4209541, 4209542, 4209543, 4209546, 4209597, 4209598, 4209599, 4209600, 4209601, 4209602, 4209604, 4209605, 4209606, 4209607, 4209608, 4209609, 4209610, 4209611, 4209612, 4209613, 4209614, 4209615, 4209616, 4209621, 4209628, 4209632, 4209634, 4209635, 4209637, 4209639, 4209640, 4209647, 4209667, 4209671, 4209685, 4209694, 4209698, 4209734, 4209764, 4209771, 4209774, 4209775, 4209776, 4209777, 4209778, 4209779, 4209780, 4209781, 4209782, 4209783, 4209784, 4209785, 4209786, 4209787, 4209788, 4209789, 4209790, 4209791, 4209792, 4209793, 4209794, 4209795, 4209796, 4209797, 4209798, 4209799, 4209800, 4209801, 4209802, 4209803, 4209804, 4209805, 4209806, 4209807, 4209808, 4209809, 4209810, 4209811, 4209812, 4209813, 4209814, 4209815, 4209816, 4209817, 4209818, 4209819, 4209820, 4209821, 4209822, 4209823, 4209834, 4209849, 4209859, 4209860, 4209863, 4209927, 4209928, 4209929, 4209930, 4209931, 4209932, 4209933, 4209934, 4209935, 4209936, 4209937, 4209938, 4209939, 4209940, 4209941, 4209942, 4209943, 4209945, 4209946, 4209947, 4209948, 4209949, 4209950, 4209951, 4209952, 4209954, 4209955, 4209956, 4209957, 4209959, 4209960, 4209961, 4209962, 4209963, 4209964, 4209965, 4209966, 4209967, 4209968, 4209969, 4209985, 4209986, 4209987, 4209988 Updated information submitted: Serial numbers removed from 12/23/20 letter: 4200112, 4200117, 4200202, 4201412, 4201490, 4201661; reason: These devices are still in scope and exhibit the issue. However, it was found that none of these devices were distributed to a customer. Therefore, they were removed from the attachment in the letter and are within Stryker control. Serial Numbers added to the recall letter: 4209292, 4209646, 4209841, 4202629, 4203963; reason: These devices were originally thought to be in Stryker control, but were shipped to customers. Serial Numbers added to the recall letter: 4209917, 4209922, 4209924; reason: These devices were shipped to Stryker OUS distributor and were thought to be distributed to customers. (Firm later confirmed that the units are under Stryker control.)
Quantity recalled367 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0472-2021
FDA event ID
86700
Recalling firm
Mako Surgical, Davie, FL
Classification
Class II
Status
Completed
Recall initiated
23 Oct 2020
FDA classified
19 Nov 2020
Reported
25 Nov 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

23 Oct 2020Recall initiated by company
19 Nov 2020Classified by FDA+27 days
25 Nov 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Mako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later).

Where it was distributed

Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK

Nationwide

Questions about this recall

Is Mako Integrated Cutting System (MICS) Handpiece recalled?

Yes. Mako Surgical recalled Mako Integrated Cutting System (MICS) Handpiece on 23 Oct 2020. The recall is Class II and its status is Completed. Recall number Z-0472-2021.

Why was it recalled?

Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.

Which lots are affected?

(01) 00848486030193 (21)<serial number>, Serial numbers: 4200052, 4200095, 4200112, 4200117, 4200119, 4200167, 4200190, 4200193, 4200202, 4200203, 4200204, 4200226, 4200233, 4200235, 4200248, 4200250, 4200257, 4200269, 4200270, 4200288, 4200293, 4200305, 4200350, 4200373, 4200384, 4200387, 4200391, 4200397, 4200398, 4200400, 4200409, 4200436, 4200457, 4200459, 4200477, 4200488, 4200512, 4200540, 4200548, 4200550, 4200558, 4200624, 4200645, 4200653, 4200666, 4200717, 4200766, 4200774, 4200785, 4200789, 4200801, 4200806, 4200807, 4200816, 4200843, 4200856, 4200868, 4200904, 4200917, 4201000, 4201040, 4201068, 4201089, 4201102, 4201103, 4201121, 4201167, 4201218, 4201236, 4201238, 4201253, 4201296, 4201302, 4201366, 4201367, 4201376, 4201385, 4201408, 4201412, 4201442, 4201452, 4201454, 4201477, 4201490, 4201630, 4201661, 4201714, 4201752, 4201829, 4201831, 4201851, 4202009, 4202020, 4202038, 4202044, 4202062, 4202063, 4202122, 4202146, 4202162, 4202175, 4202315, 4202333, 4202626, 4202631, 4202659, 4202729, 4202742, 4202882, 4202987, 4203180, 4203198, 4203289, 4203424, 4203441, 4203649, 4203709, 4203728, 4203730, 4203731, 4203736, 4203807, 4203826, 4204015, 4204164, 4204389, 4204448, 4204539, 4204553, 4204577, 4204673, 4204812, 4204934, 4205093, 4205190, 4205380, 4205426, 4206219, 4206271, 4209023, 4209025, 4209044, 4209045, 4209046, 4209065, 4209066, 4209067, 4209068, 4209071, 4209149, 4209151, 4209155, 4209156, 4209157, 4209162, 4209164, 4209166, 4209167, 4209171, 4209185, 4209189, 4209210, 4209211, 4209213, 4209220, 4209244, 4209250, 4209252, 4209255, 4209256, 4209259, 4209261, 4209262, 4209267, 4209268, 4209271, 4209298, 4209299, 4209303, 4209304, 4209306, 4209308, 4209312, 4209317, 4209318, 4209345, 4209346, 4209357, 4209360, 4209363, 4209364, 4209365, 4209387, 4209388, 4209389, 4209398, 4209441, 4209443, 4209460, 4209463, 4209464, 4209466, 4209471, 4209472, 4209473, 4209474, 4209475, 4209476, 4209477, 4209486, 4209487, 4209491, 4209493, 4209494, 4209495, 4209496, 4209500, 4209519, 4209522, 4209526, 4209529, 4209531, 4209532, 4209533, 4209535, 4209541, 4209542, 4209543, 4209546, 4209597, 4209598, 4209599, 4209600, 4209601, 4209602, 4209604, 4209605, 4209606, 4209607, 4209608, 4209609, 4209610, 4209611, 4209612, 4209613, 4209614, 4209615, 4209616, 4209621, 4209628, 4209632, 4209634, 4209635, 4209637, 4209639, 4209640, 4209647, 4209667, 4209671, 4209685, 4209694, 4209698, 4209734, 4209764, 4209771, 4209774, 4209775, 4209776, 4209777, 4209778, 4209779, 4209780, 4209781, 4209782, 4209783, 4209784, 4209785, 4209786, 4209787, 4209788, 4209789, 4209790, 4209791, 4209792, 4209793, 4209794, 4209795, 4209796, 4209797, 4209798, 4209799, 4209800, 4209801, 4209802, 4209803, 4209804, 4209805, 4209806, 4209807, 4209808, 4209809, 4209810, 4209811, 4209812, 4209813, 4209814, 4209815, 4209816, 4209817, 4209818, 4209819, 4209820, 4209821, 4209822, 4209823, 4209834, 4209849, 4209859, 4209860, 4209863, 4209927, 4209928, 4209929, 4209930, 4209931, 4209932, 4209933, 4209934, 4209935, 4209936, 4209937, 4209938, 4209939, 4209940, 4209941, 4209942, 4209943, 4209945, 4209946, 4209947, 4209948, 4209949, 4209950, 4209951, 4209952, 4209954, 4209955, 4209956, 4209957, 4209959, 4209960, 4209961, 4209962, 4209963, 4209964, 4209965, 4209966, 4209967, 4209968, 4209969, 4209985, 4209986, 4209987, 4209988 Updated information submitted: Serial numbers removed from 12/23/20 letter: 4200112, 4200117, 4200202, 4201412, 4201490, 4201661; reason: These devices are still in scope and exhibit the issue. However, it was found that none of these devices were distributed to a customer. Therefore, they were removed from the attachment in the letter and are within Stryker control. Serial Numbers added to the recall letter: 4209292, 4209646, 4209841, 4202629, 4203963; reason: These devices were originally thought to be in Stryker control, but were shipped to customers. Serial Numbers added to the recall letter: 4209917, 4209922, 4209924; reason: These devices were shipped to Stryker OUS distributor and were thought to be distributed to customers. (Firm later confirmed that the units are under Stryker control.)

Where was it sold?

Worldwide distribution - US Nationwide distribution and the countries of Australia, Japan, Korea, Netherlands, and Chile, and France, Germany Luxembourg, and UK

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.