CLASS II Device recall · Reported 24 Jun 2020 · FDA Enforcement Report
Mammomat Revelation recalled by Siemens Medical Solutions USA
Siemens Medical Solutions USA is recalling Mammomat Revelation, distributed nationwide in the US. The recall was initiated on 7 May 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2388-2020
- FDA event ID
- 85719
- Recalling firm
- Siemens Medical Solutions USA, Malvern, PA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 7 May 2020
- FDA classified
- 16 Jun 2020
- Reported
- 24 Jun 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Imaging Equipment
Timeline
| 7 May 2020 | Recall initiated by company | |
| 16 Jun 2020 | Classified by FDA | +40 days |
| 24 Jun 2020 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
MAMMOMAT Revelation, model no. 11343300 - Product Usage: is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals.
Where it was distributed
US Nationwide distribution.
Questions about this recall
Is Mammomat Revelation recalled?
Yes. Siemens Medical Solutions USA recalled Mammomat Revelation on 7 May 2020. The recall is Class II and its status is Terminated. Recall number Z-2388-2020.
Why was it recalled?
If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.
Which lots are affected?
Serial Numbers (distributed in the US): 1026 1057 1348 1426 1027 1274 1292 1360 1324 1728 1372 1744 1036 1038 1620 1235 1751 1423 1379 1425 1342 1740 1445 1650 1058 1409 1277 1025 1253 1225 1749 1747 1422 1275 1228 1449 1707 1767 1387 1391 1206 1217 1250 1082 1357 1354 1256 1031 1339 1034 1219 1437 1289 1359 1045 1429 1408 1717 1051 ***Updated 8/18/2020*** 1775 1784 1785 1795 1797
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Medical Solutions USA
All 796| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono floor recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis pheno recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | PA |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device is recalled | Canon Medical System, USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederland B.v | Lead & Heavy Metals |