CLASS II ONGOING Device recall · Reported 21 Jan 2026 · FDA Enforcement Report

Maverick¿ Measurement Module Spectrometer PN recalled by Repligen

Repligen is recalling Maverick¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN. The recall was initiated on 10 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Quantity recalled33

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1024-2026
FDA event ID
98260
Recalling firm
Repligen, Marlborough, MA
Classification
Class II
Status
Ongoing
Recall initiated
10 Dec 2025
FDA classified
13 Jan 2026
Reported
21 Jan 2026
Type
FDA Mandated
Reason category
Device Malfunction

Timeline

10 Dec 2025Recall initiated by company
13 Jan 2026Classified by FDA+34 days
21 Jan 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Where it was distributed

U.S.

Questions about this recall

Why was it recalled?

Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

Which lots are affected?

PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496

Where was it sold?

U.S.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.