CLASS II Device recall · Reported 16 Jul 2014 · FDA Enforcement Report

Medical Breathing Circuit Filter recalled by Pall

Pall is recalling PALL Medical Breathing Circuit Filter, distributed in 15 states. The recall was initiated on 4 Jun 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF BB50T: Lots 326902, 402802, 405002; REF BB50T-BULK: Lots 228702, 229902, 328202
Quantity recalled24,650 units domestically

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2038-2014
FDA event ID
68424
Recalling firm
Pall, Port Washington, NY
Classification
Class II
Status
Terminated
Recall initiated
4 Jun 2014
FDA classified
10 Jul 2014
Reported
16 Jul 2014
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Foreign Material

Timeline

4 Jun 2014Recall initiated by company
10 Jul 2014Classified by FDA+36 days
16 Jul 2014Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Possible leak from the filter allowing a small of amount of air to be released from the junction between the filter end cap (flat side of the filter) and the filter housing connection.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

PALL Medical Breathing Circuit Filter, REF BB50T and BB50T-BULK , Rx ONLY Single use, bi-directional bacterial/viral removal filter. The filters are individually packaged in a clear, plastic pouch with black writing - 50 per case, or in bulk packaged 100 filters/bag and two bags per shipper for a total of 200 filters/shipper.

Where it was distributed

Distributed in the states of MN, MO, OH, MA, NY, NJ, WY, AZ, CA, IL, LA, IN, CO, NC, and TX.

ArizonaCaliforniaColoradoIllinoisIndianaLouisianaMassachusettsMinnesotaMissouriNorth CarolinaNew JerseyNew YorkOhioTexasWyoming

Questions about this recall

Is Medical Breathing Circuit Filter recalled?

Yes. Pall recalled Medical Breathing Circuit Filter on 4 Jun 2014. The recall is Class II and its status is Terminated. Recall number Z-2038-2014.

Why was it recalled?

Possible leak from the filter allowing a small of amount of air to be released from the junction between the filter end cap (flat side of the filter) and the filter housing connection.

Which lots are affected?

REF BB50T: Lots 326902, 402802, 405002; REF BB50T-BULK: Lots 228702, 229902, 328202

Where was it sold?

Distributed in the states of MN, MO, OH, MA, NY, NJ, WY, AZ, CA, IL, LA, IN, CO, NC, and TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.