CLASS II ONGOING Device recall · Reported 22 Oct 2025 · FDA Enforcement Report
Medical convenience kits recalled by American Contract Systems
American Contract Systems is recalling Medical convenience kits, distributed nationwide in the US. The recall was initiated on 4 Sep 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0183-2026
- FDA event ID
- 97659
- Recalling firm
- American Contract Systems, Grimes, IA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 4 Sep 2025
- FDA classified
- 16 Oct 2025
- Reported
- 22 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Spinal Implants
Timeline
| 4 Sep 2025 | Recall initiated by company | |
| 16 Oct 2025 | Classified by FDA | +42 days |
| 22 Oct 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Medical convenience kits Item Number/Description HSLM53AB SPINE PACK MILM82A DTC LAMINECTOMY PACK (PS 209475) RCNA47B NEURO ANGIO 0246971 SFLM54AB LAMINECTOMY PACK - 219659 UDLM94AA LAMINECTOMY PACK - 206064 UILM77AU LAMINECTOMY (PS 039500) UINR44P NEURO IR PERIPHERAL (PS 64405) UISH15AV SHUNT PACK (PS 039360)
Where it was distributed
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Questions about this recall
Why was it recalled?
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Which lots are affected?
ANLM40AG UDI-DI 191072232076 Lot 8192111 Bag serial number 68195596; ANPP15AR UDI-DI 191072214423 Lot 8024911 Bag serial number 68055493; HSLM53AB UDI-DI 191072224095 Lot 8061111 Bag serial number 68138767; MILM82A UDI-DI 191072231512 Lot 8361111 Bag serial number 68576542; RCNA47B UDI-DI 191072228536 Lot 8441711 Bag serial number 68680026; SFLM54AB UDI-DI 191072217400 lot 8105511 Bag serial number N/A; UDLM94AA UDI-DI 191072214089 Lot 8382011 Bag serial number 17621919; UILM77AU UDI-DI 191072210333 Lot 8056311 Bag serial number N/A; UINR44P UDI-DI 191072210340 Lot 8067511 Bag serial numbers 68138618 68140374; UISH15AV UDI-DI 191072227843 Lot 8028211 Bag serial number 68154067 * If bag serial number is listed as "N/A", then whole lot is affected
Where was it sold?
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
17 other products · FDA event 97659This product is part of a recall event; the main page for the event is Medical convenience kits recalled by American Contract Systems.
More recalls from American Contract Systems
All 213| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Convenience Kits, CV B PACK recalled by American Contract Systems | American Contract Systems | Foreign Material | Nationwide |
| Device | CLASS II | Convenience Kit, CV B PACK recalled by American Contract Systems | American Contract Systems | Foreign Material | Nationwide |
Other Spinal Implants recalls
All 1,161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |