CLASS II Device recall · Reported 19 Oct 2016 · FDA Enforcement Report

The Medix NatalCare LX Intensive Care Incubator recalled

Natus Medical is recalling The Medix NatalCare LX Intensive Care Incubator, distributed in 15 states. The recall was initiated on 25 Aug 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes5045, 5046, 5047, 5048, 5051, 5053, 5056, 5057, 5096, 5097, 5128, 5129, 5130, 5131, 5132, 5133, 5134, 5135, 5136, 5138, 5139, 5209, 5210, 5211, 5212, 5213, 5214, 5219, 5220, 5221, 5222, 5223, 5224, 5225, 5226, 5227, 5228, 5248, 5249, 5250, 5251, 5252, 5253, 5254, 5255, 5256, 5257, 5276, 5277, 5278, 5279, 5280, 5281, 5282, 5283, 5284, 5285, 5286, 5287, 5288, 5289, 8290, 5291, 5292, 5293, 5294, 5322, 5323, 5324, 5325, 5326, 5327, 5328, 5329, 5330, 5331, 5332, 5333, 5334, 5335, 5336, 5337, 5383, 5385, 5388, 5389, 5504, 5505, 5506, 5507, 5508, 5509, 5510, 5511, 5543, 5544, 5545, 5546, 5547, 5548, 5549, 5550, 5551, and 5552.
Quantity recalled104 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0056-2017
FDA event ID
74627
Recalling firm
Natus Medical, Seattle, WA
Classification
Class II
Status
Terminated
Recall initiated
25 Aug 2016
FDA classified
7 Oct 2016
Reported
19 Oct 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Aug 2016Recall initiated by company
7 Oct 2016Classified by FDA+43 days
19 Oct 2016Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX0006; and KITNC-STLX0007.

Where it was distributed

US Distribution to the states of : AL, CA, FL, IL, IN, KY, LA, MA, MS, NJ, NY, OK, PA, SC and VA, and Internationally to Pakistan.

AlabamaCaliforniaFloridaIllinoisIndianaKentuckyLouisianaMassachusettsMississippiNew JerseyNew YorkOklahomaPennsylvaniaSouth CarolinaVirginia

Questions about this recall

Why was it recalled?

NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment.

Which lots are affected?

5045, 5046, 5047, 5048, 5051, 5053, 5056, 5057, 5096, 5097, 5128, 5129, 5130, 5131, 5132, 5133, 5134, 5135, 5136, 5138, 5139, 5209, 5210, 5211, 5212, 5213, 5214, 5219, 5220, 5221, 5222, 5223, 5224, 5225, 5226, 5227, 5228, 5248, 5249, 5250, 5251, 5252, 5253, 5254, 5255, 5256, 5257, 5276, 5277, 5278, 5279, 5280, 5281, 5282, 5283, 5284, 5285, 5286, 5287, 5288, 5289, 8290, 5291, 5292, 5293, 5294, 5322, 5323, 5324, 5325, 5326, 5327, 5328, 5329, 5330, 5331, 5332, 5333, 5334, 5335, 5336, 5337, 5383, 5385, 5388, 5389, 5504, 5505, 5506, 5507, 5508, 5509, 5510, 5511, 5543, 5544, 5545, 5546, 5547, 5548, 5549, 5550, 5551, and 5552.

Where was it sold?

US Distribution to the states of : AL, CA, FL, IL, IN, KY, LA, MA, MS, NJ, NY, OK, PA, SC and VA, and Internationally to Pakistan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Natus Medical

All 17

Other Other Medical Devices recalls

All 13,778