CLASS II Device recall · Reported 19 Jul 2017 · FDA Enforcement Report

The Olympic Pasteurmatic recalled by Natus Medical Incorporated

Natus Medical is recalling The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and, distributed nationwide in the US. The recall was initiated on 22 May 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes************Serial numbers of affected units distributed to US government*************** 3056 and 9199. ************Serial numbers of affected units distributed to US consignees*************** 3039, 3041, 3043, 3045, 3050, 3055, 3062, 3071, 3079, 3083, 3084, 3096, 3100, 3113, 3143, 3153, 3155, 3162, 3165, 3174, 3188, 3192, 3194, 3195, 3223, 3224, 3229, 3232, 3238, 3240, 3249, 3260, 3265, 9197, 9198, 03003, 03004, 03012, 03020, 03022, 03028, 03034, 03047, 3002, 3006, 3009, 3013, 3029, 3035, 3044, 3046, 3047, 3048, 3052, 3059, 3063, 3084, 3085, 3091, 3091, 3094, 3099, 3101, 3110, 3118, 3134, 3136, 3142, 3145, 3148, 3149, 3156, 3161, 3169, 3171, 3173, 3180, 3181, 3183, 3185, 3189, 3190, 3198, 3225, 3227, 3237, 3239, 3241, 3266, 3277, 3278, 3279, 3281, 3345, and 9198. ************Serial numbers of affected units distributed to international consignees******* 3019, 3036, 3040, 3049, 3051, 3053, 3054, 3057, 3061, 3064, 3065, 3067, 3072, 3073, 3074, 3077, 3078, 3080, 3081, 3082, 3083, 3084, 3086, 3087, 3088, 3089, 3090, 3092, 3093, 3097, 3098, 3102, 3103, 3104, 3106, 3107, 3109, 3113, 3116, 3117, 3120, 3123, 3124, 3129, 3130, 3131, 3132, 3135, 3137, 3138, 3139, 3144, 3146, 3147, 3150, 3151, 3152, 3154, 3158, 3163, 3164, 3166, 3167, 3172, 3172, 3175, 3177, 3178, 3179, 3182, 3186, 3187, 3193, 3196, 3204, 3206, 3208, 3210, 3211, 3211, 3215, 3217, 3218, 3219, 3220, 3221, 3222, 3226, 3228, 3230, 3231, 3233, 3234, 3235, 3242, 3243, 3244, 3245, 3246, 3247, 3250, 3251, 3252, 3253, 3254, 3257, 3261, 3264, 3267, 3268, 3269, 3270, 3272, 3273, 3274, 3276, 3280, 03008, 03014, 03017, 03018, 03021, 03023, 03024, 03027, 03175, 03200, 3025, 3035, 3037, 3058, 3060, 3068, 3070, 3075, 3076, 3095, 3105, 3108, 3111, 3114, 3115, 3119, 3121, 3122, 3125, 3126, 3128, 3133, 3140, 3141, 3157, 3159, 3160, 3168, 3170, 3176, 3184, 3191, 3200, 3201, 3202, 3203, 3205, 3207, 3209, 3212, 3213, 3214, 3216, 3236, 3248, 3258, 3259, 3262, 3271, 3275, and 9200.
Quantity recalledtotal 275 units (97 units in US and 178 units international)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2701-2017
FDA event ID
76427
Recalling firm
Natus Medical, Seattle, WA
Classification
Class II
Status
Terminated
Recall initiated
22 May 2017
FDA classified
7 Jul 2017
Reported
19 Jul 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

22 May 2017Recall initiated by company
7 Jul 2017Classified by FDA+46 days
19 Jul 2017Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Olympic Pasteurmatic 3000/3500 Systems are recalled because they do not fully meet todays regulatory requirements. Given the age of the products design, component and technology obsolescence, and the time involved in fully complying with the latest design and development standards, the firm decided to recall of the Olympic Pasteurmatic 3000/3500 System devices by means of disabling the devices.

Product as listed by the FDA

The Olympic Pasteurmatic 3000 and Olympic Pasteurmatic 3500 are intended for cleaning and high-level disinfection of respiratory therapy and anesthesia equipment by killing vegetative bacteria and viruses, in a hot water bath at ~160¿ - ~170¿ F for 30 minutes. Catalog Number/Description: 007168 PM3500 ASSY (15KW 208/120V) 007169 PM3000 ASSY (9KW 208/120V) 007169R PM3 (9KW 208/120) REFURB 007170 PM3 ASSY (8.3KW 200V DELTA) 007171 PM3 ASSY (9KW 220V SING PH) 52640 PASTEURMATIC 3000 W/O BASKETS 52640R PM 3000 w/o Baskets Refurb 52640-200 PM 3000 W/O BASKETS 200V 52640-220 PM 3000 W/O BASKETS 220V 52641 PASTEURMATIC 3000 W/BASKET S 52641R PM 3000 w/ Baskets Refurb 52641-200 PM 3000 W BASKETS 200V 52641-220 PM 3000 W/ BASKETS 220V 52650 PASTEURMATIC 3500 W/O BASKET 52651 PASTEURMATIC 3500 W/BASKETS S

Where it was distributed

Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AR, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, and WV and in the following countries: Canada, Colombia, Costa Rica, Ecuador, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Mexico, Panama, Peru, Qatar, Saudi Arabia, Taiwan, Thailand, and Viet Nam.

Nationwide AlabamaArkansasCaliforniaColoradoFloridaGeorgiaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiNorth CarolinaNorth DakotaNew Hampshire

Questions about this recall

Why was it recalled?

Olympic Pasteurmatic 3000/3500 Systems are recalled because they do not fully meet todays regulatory requirements. Given the age of the products design, component and technology obsolescence, and the time involved in fully complying with the latest design and development standards, the firm decided to recall of the Olympic Pasteurmatic 3000/3500 System devices by means of disabling the devices.

Which lots are affected?

************Serial numbers of affected units distributed to US government*************** 3056 and 9199. ************Serial numbers of affected units distributed to US consignees*************** 3039, 3041, 3043, 3045, 3050, 3055, 3062, 3071, 3079, 3083, 3084, 3096, 3100, 3113, 3143, 3153, 3155, 3162, 3165, 3174, 3188, 3192, 3194, 3195, 3223, 3224, 3229, 3232, 3238, 3240, 3249, 3260, 3265, 9197, 9198, 03003, 03004, 03012, 03020, 03022, 03028, 03034, 03047, 3002, 3006, 3009, 3013, 3029, 3035, 3044, 3046, 3047, 3048, 3052, 3059, 3063, 3084, 3085, 3091, 3091, 3094, 3099, 3101, 3110, 3118, 3134, 3136, 3142, 3145, 3148, 3149, 3156, 3161, 3169, 3171, 3173, 3180, 3181, 3183, 3185, 3189, 3190, 3198, 3225, 3227, 3237, 3239, 3241, 3266, 3277, 3278, 3279, 3281, 3345, and 9198. ************Serial numbers of affected units distributed to international consignees******* 3019, 3036, 3040, 3049, 3051, 3053, 3054, 3057, 3061, 3064, 3065, 3067, 3072, 3073, 3074, 3077, 3078, 3080, 3081, 3082, 3083, 3084, 3086, 3087, 3088, 3089, 3090, 3092, 3093, 3097, 3098, 3102, 3103, 3104, 3106, 3107, 3109, 3113, 3116, 3117, 3120, 3123, 3124, 3129, 3130, 3131, 3132, 3135, 3137, 3138, 3139, 3144, 3146, 3147, 3150, 3151, 3152, 3154, 3158, 3163, 3164, 3166, 3167, 3172, 3172, 3175, 3177, 3178, 3179, 3182, 3186, 3187, 3193, 3196, 3204, 3206, 3208, 3210, 3211, 3211, 3215, 3217, 3218, 3219, 3220, 3221, 3222, 3226, 3228, 3230, 3231, 3233, 3234, 3235, 3242, 3243, 3244, 3245, 3246, 3247, 3250, 3251, 3252, 3253, 3254, 3257, 3261, 3264, 3267, 3268, 3269, 3270, 3272, 3273, 3274, 3276, 3280, 03008, 03014, 03017, 03018, 03021, 03023, 03024, 03027, 03175, 03200, 3025, 3035, 3037, 3058, 3060, 3068, 3070, 3075, 3076, 3095, 3105, 3108, 3111, 3114, 3115, 3119, 3121, 3122, 3125, 3126, 3128, 3133, 3140, 3141, 3157, 3159, 3160, 3168, 3170, 3176, 3184, 3191, 3200, 3201, 3202, 3203, 3205, 3207, 3209, 3212, 3213, 3214, 3216, 3236, 3248, 3258, 3259, 3262, 3271, 3275, and 9200.

Where was it sold?

Worldwide Distribution - USA (nationwide) Distribution to the states of : AL, AR, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, and WV and in the following countries: Canada, Colombia, Costa Rica, Ecuador, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Mexico, Panama, Peru, Qatar, Saudi Arabia, Taiwan, Thailand, and Viet Nam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Natus Medical

All 17

Other CPAP & Ventilators recalls

All 554