CLASS II ONGOING Device recall · Reported 2 Sep 2026 · FDA Enforcement Report
Medline breathing circuits labeled as: Hudson Rci, Circuit recalled
Medline Industries is recalling Medline breathing circuits labeled as: Hudson Rci, Circuit, Htd, DL, Neonate, 48, Medline, distributed nationwide in the US. The recall was initiated on 13 Jul 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2992-2026
- FDA event ID
- 99403
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Jul 2026
- FDA classified
- 21 Aug 2026
- Reported
- 2 Sep 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 13 Jul 2026 | Recall initiated by company | |
| 21 Aug 2026 | Classified by FDA | +39 days |
| 2 Sep 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).
Where it was distributed
Worldwide distribution - US Nationwide and the country of Panama.
Questions about this recall
Why was it recalled?
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Which lots are affected?
1) Medline Product Number/SKU HUD78007, UDI/DI 10197344042463 (each) 20197344042460 (case), All lot numbers; 2) Medline Product Number/SKU HUD78009, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 3) Medline Product Number/SKU 780-13, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 4) Medline Product Number/SKU 780-14, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 5) Medline Product Number/SKU 780-15, UDI/DI Discontinued (each) Discontinued (case), All lot numbers; 6) Medline Product Number/SKU HUD78018, UDI/DI 10197344010660 (each) 20197344010667 (case), All lot numbers.
Where was it sold?
Worldwide distribution - US Nationwide and the country of Panama.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 99403| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide | |
| CLASS II | Medline breathing circuits labeled as: Medline Hudson Rci recalled | Medline Industries | Device Malfunction | Nationwide |
More recalls from Medline Industries
All 707| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Peds Sinus Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Mfc Craniotomy Pack recalled | Medline Industries | Other | Nationwide |
Other Other Medical Devices recalls
All 13,884| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | NOxBOXi Nitric Oxide Delivery System recalled | Noxbox | Device Malfunction | Nationwide |
| Device | CLASS II | Palacos R pro recalled by Heraeus Medical GmbH (Dental | Heraeus Medical GmbH (Dental Division) | Packaging Defects | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065 recalled | Percussionaire | Other | Nationwide |
| Device | CLASS II | Positron Emission Tomography and Computed Tomography System recalled | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Intrapulmonary Percussive Ventilation, IPV-1C recalled by Percussionaire | Percussionaire | Other | Nationwide |