CLASS II ONGOING Device recall · Reported 24 Jun 2026 · FDA Enforcement Report

Medline convenience kits: Breast Imaging DYNJT5893 Chog recalled

Medline Industries is recalling Medline convenience kits: Breast Imaging DYNJT5893 Chog Plastics SYNJ10306A Endo Breast, distributed nationwide in the US. The recall was initiated on 27 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesDYNJT5893 UDI-DI 10198459517839 Lot 25KBA123 SYNJ10306A UDI-DI 10193489296891 Lots 25LBO305 26CBD081 DYNJ68023 UDI-DI 10193489468083 Lot 25KBI169 DYNJ909926C UDI-DI 10198459423109 Lots 25LBC067 25LBR659 DYNJ84081B UDI-DI 10198459462474 Lot 25LMI881 DYNJ905362C UDI-DI 10198459346842 Lots 25KBN239 25LBF694 DYNJ0660040R UDI-DI 10889942220083 Lots 25JME572 26AMA936 DYNJ89023A UDI-DI 10198459388316 Lot 25KBC500 DYNJ902510M UDI-DI 10198459683374 Lots 26CBM279 26CBO942 DYNJ84766 UDI-DI 10195327414108 Lot 26BBC049 DYNJ0578842M UDI-DI 10193489288292 Lot 25LMG027 DYNJ63803C UDI-DI 10198459149825 Lots 25LBN476 26CBC156 DYNJ62433D UDI-DI 10198459498176 Lots 25KBC752 25KBM917 DYNJ81772C UDI-DI 10198459349362 Lots 25LMH594 25LMI396 DYNJT7588 UDI-DI 10198459635182 Lot 26AMA250 DYNJ89048A UDI-DI 10198459388477 Lot 26BBD124
Quantity recalled49,654 kits total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2423-2026
FDA event ID
98951
Recalling firm
Medline Industries, Northfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
27 Apr 2026
FDA classified
12 Jun 2026
Reported
24 Jun 2026
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Device Malfunction

Timeline

27 Apr 2026Recall initiated by company
12 Jun 2026Classified by FDA+46 days
24 Jun 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Medline convenience kits: BREAST IMAGING DYNJT5893 CHOG PLASTICS SYNJ10306A ENDO BREAST AUGMENTATION PACK DYNJ68023 FACIAL PLASTIC DYNJ909926C JVL - MODULE PLASTIC PK DYNJ84081B KIT MAJOR PLASTICS MOSES DYNJ905362C MEM MAXILLOFACIAL PACK-LF DYNJ0660040R MINOR BREAST PACK DYNJ89023A PLASTIC MAJOR SURGERY DYNJ902510M PLASTIC PACK DYNJ84766 PLASTIC PACK-LF DYNJ0578842M PLASTICS EXTREMITY PACK DYNJ63803C PLASTICS PK - NO SYR DYNJ62433D RR-ASC-RHINOPLASTY PACK-LF DYNJ81772C TRANSGENDER PACK DYNJT7588 WH PLASTIC MAJOR PACK DYNJ89048A

Where it was distributed

US Nationwide distribution. OUS distribution pending.

Nationwide

Questions about this recall

Why was it recalled?

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Which lots are affected?

DYNJT5893 UDI-DI 10198459517839 Lot 25KBA123 SYNJ10306A UDI-DI 10193489296891 Lots 25LBO305 26CBD081 DYNJ68023 UDI-DI 10193489468083 Lot 25KBI169 DYNJ909926C UDI-DI 10198459423109 Lots 25LBC067 25LBR659 DYNJ84081B UDI-DI 10198459462474 Lot 25LMI881 DYNJ905362C UDI-DI 10198459346842 Lots 25KBN239 25LBF694 DYNJ0660040R UDI-DI 10889942220083 Lots 25JME572 26AMA936 DYNJ89023A UDI-DI 10198459388316 Lot 25KBC500 DYNJ902510M UDI-DI 10198459683374 Lots 26CBM279 26CBO942 DYNJ84766 UDI-DI 10195327414108 Lot 26BBC049 DYNJ0578842M UDI-DI 10193489288292 Lot 25LMG027 DYNJ63803C UDI-DI 10198459149825 Lots 25LBN476 26CBC156 DYNJ62433D UDI-DI 10198459498176 Lots 25KBC752 25KBM917 DYNJ81772C UDI-DI 10198459349362 Lots 25LMH594 25LMI396 DYNJT7588 UDI-DI 10198459635182 Lot 26AMA250 DYNJ89048A UDI-DI 10198459388477 Lot 26BBD124

Where was it sold?

US Nationwide distribution. OUS distribution pending.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

33 other products · FDA event 98951

This product is part of a recall event; the main page for the event is Medline convenience kits: Cmh Central Line Dressing Change recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIMedline convenience kits: Cmh Central Line Dressing Change recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Davinci Main DYNJ908031D recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Acd DYNJ904999J Acdf DYNJ907336B recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Davinci DYNJ905836D Davinci-lf recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIBlock Tray Dynjra2119a Indwelling Nerve Block Dynjra1011a recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: WT Lavh Pack DYNJ46609F recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: Smsc-pelviscopy Pack-lf DYNJ0161619M recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: Hernia Repair Tote DYNJ908337D recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMinor Cath Lab Pack Dynj84775a Or Vascular Angio Pack recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: Labor & Delivery C-section DYNJ907895F recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline EBSI Kit EBSI1534 recalled by Medline IndustriesMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: C-section Kit DYNJ904305J recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Driveline Management Kit EBSI1688 recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Breast-hernia-port Cds-lf CDS984853L recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline CCI Nerve Block Tray DYNJRA2742 Medline CPNB DYNJRA2716 recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Add-a-cath Arterial Insertion recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Bib AV Fistula Pack DYNJ58304B recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Hysterectomy Pack-lf DYNJ20485L recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Inpatient Central LN Drsg Kit recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: C Section CDS985505G C Section recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Basic DYNJ905869G Basic Pack recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: IV Start Kit W/Hubguard DYNDV2662 recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Femoral Block Tray DYNJRA1739B recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IICenturion convenience kits: Central Line Dressing SV43 Fcp recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Abd Pack DYNJ81673B Bariatric recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: Bhs Neuro DYNJ904900N Medtrx Neuro recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Acl Pack DYNJ64161C Arthroscopy recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Discectomy Pack DYNJ86476B Ear recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kit: Kit: Dressing Change Ped DYNDC3359 recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Basic Eye DYNJ905154I Tubal recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: CVC Insertion KIT ECVC6805B recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Loop Pack DYNJ61460C Open Heart recalledMedline IndustriesDevice MalfunctionNationwide
CLASS IIMedline convenience kits: Port A Cath DYNJ906818A recalledMedline IndustriesDevice MalfunctionNationwide

More recalls from Medline Industries

All 705

Other Imaging Equipment recalls

All 2,341