CLASS II ONGOING Device recall · Reported 21 Jan 2026 · FDA Enforcement Report

Medline medical procedure convenience kits labeled as recalled

Medline Industries is recalling Medline medical procedure convenience kits labeled as: Kit Robotics Urology Prostate, distributed nationwide in the US. The recall was initiated on 16 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426
Quantity recalled12 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1090-2026
FDA event ID
98197
Recalling firm
Medline Industries, Northfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
16 Dec 2025
FDA classified
14 Jan 2026
Reported
21 Jan 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

16 Dec 2025Recall initiated by company
14 Jan 2026Classified by FDA+29 days
21 Jan 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Which lots are affected?

UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 98197

More recalls from Medline Industries

All 709

Other Other Medical Devices recalls

All 13,908