CLASS II Device recall · Reported 25 Dec 2019 · FDA Enforcement Report

Medtronic Adapter Percutaneous Pin recalled by Medtronic Navigation

Medtronic Navigation is recalling Medtronic Adapter Percutaneous Pin, distributed nationwide in the US. The recall was initiated on 6 Dec 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPercutaneous pin adapter lot numbers: 150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.
Quantity recalled1,297 percutaneous pin adapters

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0707-2020
FDA event ID
84045
Recalling firm
Medtronic Navigation, Louisville, CO
Classification
Class II
Status
Terminated
Recall initiated
6 Dec 2018
FDA classified
19 Dec 2019
Reported
25 Dec 2019
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

6 Dec 2018Recall initiated by company
19 Dec 2019Classified by FDA+378 days
25 Dec 2019Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.

Product as listed by the FDA

Medtronic Adapter Percutaneous Pin, REF 9734752 (GTIN 00643169529489) and REF 9734752-G02 (GTIN 00763000054915), contained in the following kits: Upgrade Set Percutaneous Pin, REF 9735502 (GTIN 00643169529533) and REF 9735502-G02 (GTIN 00763000208462); Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169540521); and Medtronic Adapter Percutaneous Pin Kit, REF 9734752K (GTIN 00643169641006) (Japan). The percutaneous pin adapter is a component of the StealthAir Frame Assembly.

Where it was distributed

Distribution was nationwide. There was government/military/foreign distribution.

Nationwide

Questions about this recall

Why was it recalled?

Under certain circumstances, the percutaneous pin adapter used with the StealthAir Frame Assembly may slightly rotate after surgical placement and not return to its original position, even when connections are tight and secure.

Which lots are affected?

Percutaneous pin adapter lot numbers: 150114, 150505, 150506, 150904, 151005, 151007 160201, 160224, 160419, 160617, 160707, 170216, 180110, and 180409.

Where was it sold?

Distribution was nationwide. There was government/military/foreign distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medtronic Navigation

All 92

Other Other Medical Devices recalls

All 13,908