CLASS II Device recall · Reported 3 Oct 2018 · FDA Enforcement Report

Medtronic Clip Gun Magazines recalled by Medtronic Neurosurgery

Medtronic Neurosurgery is recalling Medtronic Clip Gun Magazines are, distributed in 38 states. The recall was initiated on 3 May 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Number/CFN: CM8902 UDI-Device Identifier (GTIN/UPN): 20643169647200 Lot Numbers: 54984; 57188; 59558; 66312; and 67988
Quantity recalled2616

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3212-2018
FDA event ID
80657
Recalling firm
Medtronic Neurosurgery, Goleta, CA
Classification
Class II
Status
Terminated
Recall initiated
3 May 2018
FDA classified
21 Sep 2018
Reported
3 Oct 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

3 May 2018Recall initiated by company
21 Sep 2018Classified by FDA+141 days
3 Oct 2018Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Medtronic Clip Gun Magazines are packaged in individually sealed pouches and distributed in boxes containing 5 individually pouched Clip Gun Magazine. REF CM8902, Medtronic Clip Gun Magazine, Rx only, CE 0344, STERILE EO, Rx only, Single Use

Where it was distributed

Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa

AlabamaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaFloridaGeorgiaIowaIdahoIllinoisIndianaKansasLouisianaMarylandMichiganMinnesotaMissouriMississippiNorth Carolina

Questions about this recall

Why was it recalled?

The packaging sterile barrier does not comply with the applicable packaging standard for terminally sterilized medical devices. While the Instructions for Use (IFU) state Do not use if package has been previously opened or damaged , it is possible that the seal breach may go undetected by the user's naked eye.

Which lots are affected?

Product Number/CFN: CM8902 UDI-Device Identifier (GTIN/UPN): 20643169647200 Lot Numbers: 54984; 57188; 59558; 66312; and 67988

Where was it sold?

Domestic: AL, AZ, AR, CA, CO, DC, FL, GA, ID, IL, IN, IA, KS, LA, MD, MI, MN, MS, MO, NB, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY OUS: Australia, Germany, Kenya, Russian Federation, South Africa

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.