CLASS II Device recall · Reported 11 Jul 2018 · FDA Enforcement Report
microTargeting" Drive System recalled by FHC
FHC is recalling microTargeting" Drive System, distributed in 46 states. The recall was initiated on 10 May 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2356-2018
- FDA event ID
- 80148
- Recalling firm
- FHC, Bowdoin, ME
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 10 May 2018
- FDA classified
- 5 Jul 2018
- Reported
- 11 Jul 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Other Medical Devices
Timeline
| 10 May 2018 | Recall initiated by company | |
| 5 Jul 2018 | Classified by FDA | +56 days |
| 11 Jul 2018 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
microTargeting" Drive System Distributed by Medtronic, Product Number 9033G0601
Where it was distributed
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
AlabamaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaine
Questions about this recall
Is microTargeting" Drive System recalled?
Yes. FHC recalled microTargeting" Drive System on 10 May 2018. The recall is Class II and its status is Terminated. Recall number Z-2356-2018.
Why was it recalled?
The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Which lots are affected?
REF 9033G0601
Where was it sold?
The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from FHC
All 17| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | microTargeting Insertion Tube Set recalled by FHC | FHC | Other | 6 states |
| Device | CLASS II | Guideline 4000 0 GL5 Main Processing Unit recalled by FHC | FHC | Other | Nationwide |
| Device | CLASS II | microTargeting" STar" Drive System (Motorized) recalled by FHC | FHC | Lead & Heavy Metals | 46 states |
| Device | CLASS II | microTargeting" STar" Drive System (Motorized) recalled by FHC | FHC | Lead & Heavy Metals | 46 states |
| Device | CLASS II | microTargeting" Drive System recalled by FHC | FHC | Lead & Heavy Metals | 46 states |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |