CLASS II Device recall · Reported 11 Jul 2018 · FDA Enforcement Report

microTargeting" STar" Drive System (Motorized) recalled by FHC

FHC is recalling microTargeting" STar" Drive System (Motorized), Product Number 70-Zd-me, distributed in 46 states. The recall was initiated on 10 May 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 00873263002155
Quantity recalled2952 total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2352-2018
FDA event ID
80148
Recalling firm
FHC, Bowdoin, ME
Classification
Class II
Status
Terminated
Recall initiated
10 May 2018
FDA classified
5 Jul 2018
Reported
11 Jul 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

10 May 2018Recall initiated by company
5 Jul 2018Classified by FDA+56 days
11 Jul 2018Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Product as listed by the FDA

microTargeting" STar" Drive System (Motorized), Product Number 70-ZD-ME

Where it was distributed

The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.

AlabamaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaine

Questions about this recall

Why was it recalled?

The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.

Which lots are affected?

UDI: 00873263002155

Where was it sold?

The devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign countries: Argentina, Australia, Canada, Columbia, Denmark, Ecuador, France, Germany, Great Britain, Italy, Japan, Norway, Romania, South Africa, Switzerland, Saudi Arabia, Taiwan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

14 other products · FDA event 80148

This product is part of a recall event; the main page for the event is microTargeting" STar" Drive System (Manual), Product Number St-ds-ma recalled.

More recalls from FHC

All 17
ReportedClassRecallCompanyReasonWhere
DeviceCLASS ImicroTargeting Insertion Tube Set recalled by FHCFHCOther6 states
DeviceCLASS IIGuideline 4000 5.0 GL5 Main Processing Unit recalled by FHCFHCOtherNationwide
DeviceCLASS IImicroTargeting" STar" Drive System (Motorized) recalled by FHCFHCOther46 states
DeviceCLASS IImicroTargeting" Drive System recalled by FHCFHCOther46 states
DeviceCLASS IImicroTargeting" Drive System, Product Number 66-Zd-md recalled by FHCFHCOther46 states

Other Other Medical Devices recalls

All 13,908