CLASS II ONGOING Device recall · Reported 1 Jan 2025 · FDA Enforcement Report

Misha Knee System Implant Small, Left recalled by Moximed

Moximed is recalling Misha Knee System Implant Small, Left, distributed nationwide in the US. The recall was initiated on 2 Dec 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI/Lot(Expiration): 2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).
Quantity recalled115

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0762-2025
FDA event ID
95887
Recalling firm
Moximed, Fremont, CA
Classification
Class II
Status
Ongoing
Recall initiated
2 Dec 2024
FDA classified
23 Dec 2024
Reported
1 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

2 Dec 2024Recall initiated by company
23 Dec 2024Classified by FDA+21 days
1 Jan 2025Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.

Where it was distributed

US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.

Nationwide CaliforniaColoradoNew YorkOhioOregonTexasVirginia

Questions about this recall

Why was it recalled?

Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.

Which lots are affected?

REF/UDI-DI/Lot(Expiration): 2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).

Where was it sold?

US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other Hip & Knee Implants recalls

All 3,111