CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report

Unity Knee Asc Modular Handle included recalled by Corin Medical

Corin Medical is recalling Unity Knee Asc Modular Handle included in the Corin Unity Asc Instrument System, distributed nationwide in the US. The recall was initiated on 2 Oct 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Quantity recalled22 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0356-2027
FDA event ID
99959
Recalling firm
Corin Medical, Cirencester
Classification
Class II
Status
Ongoing
Recall initiated
2 Oct 2026
FDA classified
3 Nov 2026
Reported
11 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

2 Oct 2026Recall initiated by company
3 Nov 2026Classified by FDA+32 days
11 Nov 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Where it was distributed

Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.

Nationwide CaliforniaGeorgiaIllinoisMichiganNorth CarolinaRhode IslandTexasUtah

Questions about this recall

Why was it recalled?

Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.

Which lots are affected?

Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.

Where was it sold?

Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Corin Medical

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIApex Knee System recalled by Corin MedicalCorin MedicalPackaging DefectsNationwide

Other Hip & Knee Implants recalls

All 3,123