CLASS II Device recall · Reported 20 Feb 2013 · FDA Enforcement Report
Mizuho Orthopedic Table recalled by Mizuho OSI
Mizuho OSI is recalling Mizuho Orthopedic Table, distributed in 48 states. The recall was initiated on 17 Dec 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0810-2013
- FDA event ID
- 63991
- Recalling firm
- Mizuho OSI, Union City, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Dec 2012
- FDA classified
- 13 Feb 2013
- Reported
- 20 Feb 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Hip & Knee Implants
Timeline
| 17 Dec 2012 | Recall initiated by company | |
| 13 Feb 2013 | Classified by FDA | +58 days |
| 20 Feb 2013 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.
Where it was distributed
USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Is Mizuho Orthopedic Table recalled?
Yes. Mizuho OSI recalled Mizuho Orthopedic Table on 17 Dec 2012. The recall is Class II and its status is Terminated. Recall number Z-0810-2013.
Why was it recalled?
Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.
Which lots are affected?
Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.
Where was it sold?
USA Distribution -- Worldwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Mizuho OSI
All 6| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Hana/PROFx Classic Femoral Hook Right recalled by Mizuho OSI | Mizuho OSI | Other | Nationwide |
| Device | CLASS II | Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA recalled | Mizuho OSI | Other | Nationwide |
| Device | CLASS II | Levo Arm Label: LEVO ARM recalled by Mizuho OSI | Mizuho OSI | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | 216 Wilson Frame Pads recalled by Mizuho OSI | Mizuho OSI | Foreign Material | CA |
| Device | CLASS II | Trios, Table Base recalled by Mizuho OSI | Mizuho OSI | Other | CA |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Nitrosamine Impurity | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Lead & Heavy Metals | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |