CLASS II ONGOING Device recall · Reported 2 Jul 2025 · FDA Enforcement Report
Model: Description/REF: Mna-420: Mna-420 ENC -H/K9303320 recalled
Barco N.V is recalling Model: Description/REF: Mna-420: Mna-420 ENC -H/K9303320, Mna-420 DEC, distributed nationwide in the US. The recall was initiated on 9 May 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1994-2025
- FDA event ID
- 96885
- Recalling firm
- Barco N.V, Kortrijk
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 9 May 2025
- FDA classified
- 20 Jun 2025
- Reported
- 2 Jul 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 9 May 2025 | Recall initiated by company | |
| 20 Jun 2025 | Classified by FDA | +42 days |
| 2 Jul 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
Product as listed by the FDA
Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
Where it was distributed
US Nationwide distribution in the states of TX, OH.
Questions about this recall
Why was it recalled?
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
Which lots are affected?
REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726; K9303300/05415334002180/2531537815, 2531557845, 2531557866, 2531558588, 2531558727, 2531558738, 2531558753, 2531558754, 2531558759, 2531561761, 2531561785, 2531572541, 2531572557, 2531572568, 2531572576, 2531572847, 2531572862, 2531572872, 2531633606, 2531633662; K9303311/05415334026810/2531537674, 2531572141, 2531580810, 2531580935, 2531582319, 2531582331, 2531582341, 2531588386, 2531588401, 2531588404, 2531588456, 2531591560, 2531592036, 2531592037, 2531592046, 2531593872, 2531594457, 2531594475, 2531594481, 2531594503, 2531594515, 2531597664, 2531597713, 2531597717, 2531597726, 2531597795, 2531597806, 2531597809, 2531597810, 2531597820, 2531599836, 2531605356, 2531605390; K9306210/2531542260, 2531542263, 2531542270, 2531542292, 2531543553, 2531543568, 2531543588, 2531582676, 2531582691, 2531588558, 2531588562, 2531593741, 2531593749, 2531593752, 2531593763, 2531593764, 2531593768, 2531593777, 2531593784, 2531595829, 2531595831, 2531601786; K9306220/ 2531540107, 2531542441, 2531542470, 2531543538, 2531543549, 2531543611, 2531543613, 2531543640, 2531563132, 2531572740, 2531572759, 2531572770, 2531572814, 2531572827, 2531575652, 2531594331, 2531594332, 2531594336, 2531594348, 2531594350; K9306410/2531572228, 2531582713, 2531582724, 2531582725, 2531584168, 2531584176, 2531584191, 2531588591, 2531588593, 2531588598, 2531588616, 2531588626, 2531604519, 2531604535, 2531604544, 2531604545, 2531604549, 2531604550, 2531604551, 2531604593, 2531604637, 2531604649, 2531604658; K9306420/ 2531537844, 2531537870, 2531537889, 2531537901, 2531537902, 2531556037, 2531556043, 2531572419, 2531572433, 2531572440, 2531595759
Where was it sold?
US Nationwide distribution in the states of TX, OH.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Barco N.V
All 5| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Mna-420 Enc Hdmi, Material Number K9303311 recalled by Barco N.V | Barco N.V | Other | FL |
| Device | CLASS II | Barco Mdsc-8527 NXF recalled by Barco N.V | Barco N.V | Other | Nationwide |
| Device | CLASS II | Barco Mdnc-3421 Display, Diagnostic Radiology recalled by Barco N.V | Barco N.V | Other | WI |
| Device | CLASS II | Optional Mdrc-1119 Front Cover Attachment Kit recalled by Barco N.V | Barco N.V | Foreign Material | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |