CLASS II ONGOING Device recall · Reported 8 Sep 2021 · FDA Enforcement Report
Monaco RTP System recalled by Elekta
Elekta is recalling Monaco RTP System, distributed in 7 states. The recall was initiated on 26 Jul 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2406-2021
- FDA event ID
- 88451
- Recalling firm
- Elekta, Saint Charles, MO
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 26 Jul 2021
- FDA classified
- 2 Sep 2021
- Reported
- 8 Sep 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Medical Software
Timeline
| 26 Jul 2021 | Recall initiated by company | |
| 2 Sep 2021 | Classified by FDA | +38 days |
| 8 Sep 2021 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10
Where it was distributed
Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom
Questions about this recall
Why was it recalled?
If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.
Which lots are affected?
Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10 UDI: (01)00858164002190(10) 5.40.00 (01)00858164002190(10) 5.40.01 (01)00858164002190(10) 5.40.02 (01)00858164002268(10) 5.51.10
Where was it sold?
Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Elekta
All 122| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Geneva Brachytherapy Applicator recalled by Elekta | Elekta | Other | Nationwide |
| Device | CLASS II | Brachytherapy Applicator Advanced Gynecological Applicator recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Fletcher CT/MR Shielded Applicator Set Intracavitary Remote recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Mosaiq Oncology Information System recalled by Elekta | Elekta | Device Malfunction | Nationwide |
| Device | CLASS II | Leksell Vantage Arc System recalled by Elekta | Elekta | Device Malfunction | 28 states |
Other Medical Software recalls
All 938| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | intelerad IntelePACS recalled by Intelerad Medical Systems | Intelerad Medical Systems | Device Malfunction | 36 states |
| Device | CLASS I | Prosthesis, mitral valve, percutaneously delivered Component: No recalled | Edwards Lifesciences | Device Malfunction | 24 states |
| Device | CLASS II | Nalu TD2 Therapy Disc with firmware recalled by Boston Scientific | Boston Scientific Neuromodulation | Other | Nationwide |
| Device | CLASS II | Metatarsophalangeal (MTP) Implant is a sterile recalled by restor3d | restor3d | Sterility | Nationwide |
| Device | CLASS II | Ion Torrent Genexus Dx recalled by Life Technologies Holdings Pte | Life Technologies Holdings Pte | Device Malfunction | Nationwide |