CLASS II ONGOING Device recall · Reported 25 Feb 2026 · FDA Enforcement Report
Monarch Platform, Bronchoscopy recalled by Auris Health
Auris Health is recalling Monarch Platform, Bronchoscopy, distributed nationwide in the US. The recall was initiated on 21 Jan 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1383-2026
- FDA event ID
- 98270
- Recalling firm
- Auris Health, Santa Clara, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 21 Jan 2026
- FDA classified
- 17 Feb 2026
- Reported
- 25 Feb 2026
- Type
- Voluntary: Firm initiated
- Reason category
- Lead & Heavy Metals
- Device type
- Medical Software
Timeline
| 21 Jan 2026 | Recall initiated by company | |
| 17 Feb 2026 | Classified by FDA | +27 days |
| 25 Feb 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Where it was distributed
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Nationwide AlabamaArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIdahoIllinoisIndianaLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiMontana
Questions about this recall
Is Monarch Platform, Bronchoscopy recalled?
Yes. Auris Health recalled Monarch Platform, Bronchoscopy on 21 Jan 2026. The recall is Class II and its status is Ongoing. Recall number Z-1383-2026.
Why was it recalled?
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Which lots are affected?
Software versions: 2.2.2, 2.2.3, 2.2.4, 2.2.5, and 4.1.1. UDI-DI: 10810068810803, 10810068810988. Serial Numbers: 110024, 120088, 120057, 120046, 120058, 120015, 120160, 120068, 120081, 120056, 110028, 120128, 120017, 120089, 120075, 110016, 120031, 120169, 120113, 120168, 120055, 120014, 120060, 120174, 120195, 120026, 110038, 120124, 120194, 110027, 110035, 120123, 120091, 120007, 110022, 120006, 120152, 120004, 120120, 120041, 120106, 110034, 120126, 120082, 120151, 120051, 120020, 120153, 120193, 120170, 120108, 120037, 120074, 120199, 120130, 120095, 110026, 120096, 120009, 120111, 120022, 120054, 120163, 110045, 120027, 110039, 120090, 120141, 120072, 120155, 110049, 120143, 110047, 120125, 120052, 110052, 120100, 120050, 120098, 120045, 120131, 110020, 110030, 110032, 110021, 120013, 120105, 120136, 120192, 110017, 120029, 120139, 110031, 120137, 120053, 110029, 120028, 120034, 120085, 120019, 120049, 120209, 120144, 120161, 120001, 110033, 120145, 120127, 110046, 110043, 120011, 120070, 120076, 120134, 120117, 110025, 120077, 120149, 120191, 120102, 120002, 120156, 120010, 120171, 120003, 120023, 120146, 120188, 120043, 120071, 120042, 120196, 110051, 120065, 120172, 120116, 120150, 120066, 120166, 120033, 110013, 120079, 120008, 120086, 120032, 120073, 1806P1, 120197, 120080, 120078, 14, 120012, 110041, 120084, 120219, 110036, 120181, 120083, 120067, 120021, 120216, 120217, 120218, 120069, 120204, 120207, 120215, 120237, 120208, 120213, 120164, 120132, 120182
Where was it sold?
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Auris Health
All 8| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Monarch Bronchoscope recalled by Auris Health | Auris Health | Packaging Defects | Nationwide |
| Device | CLASS II | Monarch Platform (Urology) recalled by Auris Health | Auris Health | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Monarch Platform (Endoscopy) recalled by Auris Health | Auris Health | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Monarch Platform recalled by Auris Health | Auris Health | Device Malfunction | 34 states |
| Device | CLASS II | Monarch Platform recalled by Auris Health | Auris Health | Device Malfunction | 34 states |
Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel Product recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |