CLASS II Device recall · Reported 12 Jun 2013 · FDA Enforcement Report

Monopolar Curved Scissors (a.k.a recalled by Intuitive Surgical

Intuitive Surgical is recalling Monopolar Curved Scissors (a.k.a, distributed nationwide in the US. The recall was initiated on 8 May 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers 420179-09 and -10 and 400179-09 and -10.
Quantity recalled~132,441 instruments

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1442-2013
FDA event ID
65216
Recalling firm
Intuitive Surgical, Sunnyvale, CA
Classification
Class II
Status
Terminated
Recall initiated
8 May 2013
FDA classified
31 May 2013
Reported
12 Jun 2013
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Packaging Defects

Timeline

8 May 2013Recall initiated by company
31 May 2013Classified by FDA+23 days
12 Jun 2013Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

8 mm Monopolar Curved Scissors (a.k.a. Hot Shears) Manufactured and Distributed by: Intuitive Surgical Sunny vale, CA The EndoWrist Monopolar Curved Scissors Instrument is a multiple-use endoscopic instrument utilizing a single use, tip cover accessory, to be used in conjunction with the Intuitive Surgical Endoscopic Instrument Control System.

Where it was distributed

Worldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt, Guadeloupe, Israel, Ireland, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Philippines, Portugal, Poland, Qatar, Russia, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, United Kingdom, Saudi Arabia, Singapore, South Korea, and Venezuela.

Nationwide

Questions about this recall

Is Monopolar Curved Scissors (a.k.a recalled?

Yes. Intuitive Surgical recalled Monopolar Curved Scissors (a.k.a on 8 May 2013. The recall is Class II and its status is Terminated. Recall number Z-1442-2013.

Why was it recalled?

Intuitive Surgical has identified a potential for some units of the Monopolar curved Scissors version 09 and 10 instruments to develop very small cracks near the distal end. This may create a pathway for electrosurgical energy to leak to tissue and potentially cause thermal injury.

Which lots are affected?

Model Numbers 420179-09 and -10 and 400179-09 and -10.

Where was it sold?

Worldwide distribution - USA (nationwide) Foreign Distribution to Argentina, Austria, Australia, Belgium, Brazil, Bulgaria, Canada, China, Chile, Czech Republic, Denmark, Finland, France, Germany, Cyprus- Greece, Egypt, Guadeloupe, Israel, Ireland, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Pakistan, Philippines, Portugal, Poland, Qatar, Russia, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, United Kingdom, Saudi Arabia, Singapore, South Korea, and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Intuitive Surgical

All 256

Other Surgical Instruments recalls

All 3,809