CLASS II Device recall · Reported 23 Dec 2015 · FDA Enforcement Report
Mosaiq Oncology Information System recalled by Elekta
Elekta is recalling Mosaiq Oncology Information System, distributed nationwide in the US. The recall was initiated on 24 Nov 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0447-2016
- FDA event ID
- 72725
- Recalling firm
- Elekta, Atlanta, GA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 24 Nov 2015
- FDA classified
- 17 Dec 2015
- Reported
- 23 Dec 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Medical Software
Timeline
| 24 Nov 2015 | Recall initiated by company | |
| 17 Dec 2015 | Classified by FDA | +23 days |
| 23 Dec 2015 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Product as listed by the FDA
MOSAIQ Oncology Information System. Used to manage workflows for treatment planning and delivery.
Where it was distributed
Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland
Questions about this recall
Why was it recalled?
First Databank (FDB) is a third party product incorporated into MOSAIQ. FDB routinely modifies drugs in their database to reflect market and formulary changes. In FDB Framework v2.0, those drugs are not being included in drug-drug, drug-allergy and duplicate drug therapy interaction checking.
Which lots are affected?
Versions prior to 2.60.
Where was it sold?
Nationwide Distribution -- AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WY, & Puerto Rico.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Elekta
All 122| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Geneva Brachytherapy Applicator recalled by Elekta | Elekta | Other | Nationwide |
| Device | CLASS II | Brachytherapy Applicator Advanced Gynecological Applicator recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Fletcher CT/MR Shielded Applicator Set Intracavitary Remote recalled | Elekta | Other | Nationwide |
| Device | CLASS II | Mosaiq Oncology Information System recalled by Elekta | Elekta | Device Malfunction | Nationwide |
| Device | CLASS II | Leksell Vantage Arc System recalled by Elekta | Elekta | Device Malfunction | 28 states |
Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |