CLASS II Device recall · Reported 1 Aug 2018 · FDA Enforcement Report

MR Surgical Suite Option Product The MR Surgical Option recalled

GE Medical Systems is recalling MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE, distributed nationwide in the US. The recall was initiated on 30 Mar 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers M8074SS, M8053SS, M0074SS, M0001SS, 5265300-7, 5265300-8. Serial Number (System ID Number): 00000000000002 (602406IMR), 4295404 (4295404MR11), N/A (415UCBHMR3), N/A (720848MR4), 00000304124MR1 (904202MR8), N/A (309655IMRI), 4425007 (913588NMRW), 00000000UA0117 (616267IMRI), 00000000000003 (507284NMRV), 4311497 (MR229054), 4510086 (212636SIGNA3T), 00000000UA0116 (330344MR750W), UA0326 (614293MR9), N/A (614RMHTMR1), 4371613 (412692MR750W), N/A (214820ITABLE), 4250114 (409772MR6), 4550641 (4550641MR11), 4482037 (757388750WMR), 00000012228M63 (608263MR4IOR), 4356151 (414805IMR1), 00000299229MR5 (AH5824MR01), 00000012051M69 (AH5824SS01), 00000021158M61 (MRHM0885), 00000014131M67 (MRHM0323), 00000018480M64 (MR307144MR6), 4565719 (082427110075), 00000013254M68 (082445160014), 000000PTST0008 (082427040101), 000000PTST0007 (082427050024), 4596631 (082427310113), 4409769 (082427140133), 000000PTST0003 (CS1020MR01), DUMFMI60885001 (CS1006MR03), 00001309FMM07L (T4185506), 00000021695M62 (M4194477), 00000021695M62 (X41944020), 00000305145MR5 (UC2547MR01), 4320023 (083027822002316), 4508956 (A5381507), 00000010003M62 (EM0293), 00001219FMM0H1 (EM0105), 00000017217M61 (EM0105), 000000PTST0010 (YM3833), 00000000000004 (YM1733), 2937140 (0850270414), 4548529 (KW1008MR08), 000000PTST0006 (ZA2169MR01), 00001251FMM044 (34368MRS01), 4344097 (34418MRS01), 4605737 (N/A), 000000PTST0005 (10590MRS01), DUMFMI60885002 (10590MRS02), 4569734 (00132MRS01)
Quantity recalled54 units total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2527-2018
FDA event ID
80287
Recalling firm
GE Medical Systems, Waukesha, WI
Classification
Class II
Status
Terminated
Recall initiated
30 Mar 2018
FDA classified
23 Jul 2018
Reported
1 Aug 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

30 Mar 2018Recall initiated by company
23 Jul 2018Classified by FDA+115 days
1 Aug 2018Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.

Product as listed by the FDA

MR Surgical Suite Option Product The MR Surgical Option when integrated with the GE Signa¿ 3.0T or 1.5T magnetic resonance scanner (previously cleared K041476) can be used to produce head and whole body magnetic resonance images that are high resolution high signal-to-noise ratio, with short scan times. The Signa¿ Excite MR system with Surgical Option is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectroscopic images, and/or spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa Excite system with Surgical Option reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The GE Signa Excite system with Surgical Option can provide MR images at various stages of a surgical procedure, and provides the capability to transfer a patient to the MR system table directly from a surgical table, or patient transport system. Images can be obtained using the MR scanner body coil, or local surface coils, and tile surface coils used with the integrated system may / can accommodate sterile draping for surgical procedures.

Where it was distributed

US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.

Nationwide ArizonaCaliforniaColoradoFloridaIllinoisKansasMichiganMinnesotaNew YorkOhioPennsylvaniaTexasUtahVirginiaWisconsin

Questions about this recall

Why was it recalled?

There is a potential safety issue with the table transfer release of the MR Surgical Suite System which may cause an unexpected loss of function of the patient transfer mechanism that allows the patient to be moved from the GE MR Surgical Table to the Maquet Magnus Surgical Table.

Which lots are affected?

Model Numbers M8074SS, M8053SS, M0074SS, M0001SS, 5265300-7, 5265300-8. Serial Number (System ID Number): 00000000000002 (602406IMR), 4295404 (4295404MR11), N/A (415UCBHMR3), N/A (720848MR4), 00000304124MR1 (904202MR8), N/A (309655IMRI), 4425007 (913588NMRW), 00000000UA0117 (616267IMRI), 00000000000003 (507284NMRV), 4311497 (MR229054), 4510086 (212636SIGNA3T), 00000000UA0116 (330344MR750W), UA0326 (614293MR9), N/A (614RMHTMR1), 4371613 (412692MR750W), N/A (214820ITABLE), 4250114 (409772MR6), 4550641 (4550641MR11), 4482037 (757388750WMR), 00000012228M63 (608263MR4IOR), 4356151 (414805IMR1), 00000299229MR5 (AH5824MR01), 00000012051M69 (AH5824SS01), 00000021158M61 (MRHM0885), 00000014131M67 (MRHM0323), 00000018480M64 (MR307144MR6), 4565719 (082427110075), 00000013254M68 (082445160014), 000000PTST0008 (082427040101), 000000PTST0007 (082427050024), 4596631 (082427310113), 4409769 (082427140133), 000000PTST0003 (CS1020MR01), DUMFMI60885001 (CS1006MR03), 00001309FMM07L (T4185506), 00000021695M62 (M4194477), 00000021695M62 (X41944020), 00000305145MR5 (UC2547MR01), 4320023 (083027822002316), 4508956 (A5381507), 00000010003M62 (EM0293), 00001219FMM0H1 (EM0105), 00000017217M61 (EM0105), 000000PTST0010 (YM3833), 00000000000004 (YM1733), 2937140 (0850270414), 4548529 (KW1008MR08), 000000PTST0006 (ZA2169MR01), 00001251FMM044 (34368MRS01), 4344097 (34418MRS01), 4605737 (N/A), 000000PTST0005 (10590MRS01), DUMFMI60885002 (10590MRS02), 4569734 (00132MRS01)

Where was it sold?

US Nationwide distribution to AZ, CA, CO, FL, IL, KS, MI, MN, NY, OH, PA, TX, UT, VA, WI. Worldwide distribution to Austria, Brazil, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Kuwait, South Africa, Turkey, Turkey, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from GE Medical Systems

All 160

Other Spinal Implants recalls

All 1,157