CLASS II ONGOING Device recall · Reported 18 Oct 2023 · FDA Enforcement Report

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88 recalled

Techno-path Manufacturing is recalling Multichem S Plus, Part, distributed nationwide in the US. The recall was initiated on 25 Aug 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesa) 08P88-10, UDI/DI 05391523440776, Lot Numbers: 108032101, 110072101, 119112101, 126052201, 130112201, 131112201; b) 08P88-11, UDI/DI 05391523440783, Lot Numbers: 108032102, 110072102, 119112102, 126052202, 130112202, 131112202; c) 08P88-12, UDI/DI 05391523440790, Lot Numbers: 108032103, 110072103, 119112103, 126052203, 130112203, 131112203
Quantity recalled117155 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0104-2024
FDA event ID
93017
Recalling firm
Techno-path Manufacturing, Ballina Tipperary, N/A
Classification
Class II
Status
Ongoing
Recall initiated
25 Aug 2023
FDA classified
12 Oct 2023
Reported
18 Oct 2023
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Foreign Material

Timeline

25 Aug 2023Recall initiated by company
12 Oct 2023Classified by FDA+48 days
18 Oct 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Where it was distributed

US Nationwide distribution in the state of Illinois.

Nationwide Illinois

Questions about this recall

Why was it recalled?

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

Which lots are affected?

a) 08P88-10, UDI/DI 05391523440776, Lot Numbers: 108032101, 110072101, 119112101, 126052201, 130112201, 131112201; b) 08P88-11, UDI/DI 05391523440783, Lot Numbers: 108032102, 110072102, 119112102, 126052202, 130112202, 131112202; c) 08P88-12, UDI/DI 05391523440790, Lot Numbers: 108032103, 110072103, 119112103, 126052203, 130112203, 131112203

Where was it sold?

US Nationwide distribution in the state of Illinois.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 93017

This product is part of a recall event; the main page for the event is Multichem P, Part recalled by Techno-path Manufacturing.

More recalls from Techno-path Manufacturing

All 5

Other Other Medical Devices recalls