CLASS II Device recall · Reported 26 Mar 2014 · FDA Enforcement Report

Natural-Knee¿ II System: Poly Ultra Tib LT SZ00/9M Poly recalled

Zimmer is recalling Natural-Knee¿ II System: Poly Ultra Tib LT SZ00/9M Poly Ultra Tib LT SZ0/ Poly Ultra Tib, distributed nationwide in the US. The recall was initiated on 25 Feb 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem Number 621800809 621801809 621801819 621801913 621801916 621810816 621810909 621820809 621820811 621820911 621830811 621830909 621850916
Quantity recalled3638 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1262-2014
FDA event ID
67503
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
25 Feb 2014
FDA classified
20 Mar 2014
Reported
26 Mar 2014
Type
Voluntary: Firm initiated
Reason category
Sterility

Timeline

25 Feb 2014Recall initiated by company
20 Mar 2014Classified by FDA+23 days
26 Mar 2014Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Natural-Knee¿ II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY ULTRA TIB RT SZ0/13M POLY ULTRA TIB RT SZ0/16M ALL POLY ULTRA TIB LT SZ1 ALL POLY ULTRA TIB RT SZ1 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB LT SZ2 ALL POLY ULTRA TIB RT SZ2 ALL POLY ULTRA TIB LT SZ3 ALL POLY ULTRA TIB RT SZ3 ALL POLY ULTRA TIB RT SZ5

Where it was distributed

Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaIllinoisIndianaKansasLouisianaMaineMichiganMissouriMississippiNorth CarolinaNorth DakotaNebraska

Questions about this recall

Is Natural-Knee¿ II System: Poly Ultra Tib LT SZ00/9M Poly Ultra Tib LT SZ0/ Poly Ultra Tib recalled?

Yes. Zimmer recalled Natural-Knee¿ II System: Poly Ultra Tib LT SZ00/9M Poly Ultra Tib LT SZ0/ Poly Ultra Tib on 25 Feb 2014. The recall is Class II and its status is Terminated. Recall number Z-1262-2014.

Why was it recalled?

A complaint was received for a discolored All Poly Tibia. Root cause analysis found that the discoloration was the result of free radical oxidation. The oxidation process was advanced prematurely as a result of the machined component being stored in a transparent container exposed to artificial light for six years prior to inert packaging and sterilization of the device.

Which lots are affected?

Item Number 621800809 621801809 621801819 621801913 621801916 621810816 621810909 621820809 621820811 621820911 621830811 621830909 621850916

Where was it sold?

Worldwide Distribution-USA (nationwide) including DC and the states of AK, AL, AR, AZ, CA, CO, CT, FL, IL, IN, KS, LA, ME, MI, MO, MS, NC, ND, NE, NJ, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI, and the countries of Canada, Australia, India, Austria, Japan, Czech Republic, Seoul Korea, France, Venezuela, Germany, Luxembourg, Spain, and Switzerland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer

All 539

Other Hip & Knee Implants recalls

All 3,097