CLASS II Device recall · Reported 7 Sep 2016 · FDA Enforcement Report
Nio-a / Nio-p is an instant IO access device comprises recalled
Waismed is recalling Nio-a / Nio-p is an instant IO access device comprises housing with spring load mechanism, distributed in California, Louisiana, Texas, Wisconsin. The recall was initiated on 18 Aug 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2680-2016
- FDA event ID
- 75008
- Recalling firm
- Waismed, Rosh Haayin, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 18 Aug 2016
- FDA classified
- 27 Aug 2016
- Reported
- 7 Sep 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Other Reasons
- Device type
- Surgical Instruments
Timeline
| 18 Aug 2016 | Recall initiated by company | |
| 27 Aug 2016 | Classified by FDA | +9 days |
| 7 Sep 2016 | Published in enforcement report | +11 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.
Product as listed by the FDA
NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle into the bone marrow of a patient and allows connection of syringe or IV administration set. The NIO-A is intended for adults above 12 years old and the NIO-P is intended for pediatric patients 3 - 12 years of age.
Where it was distributed
WI, TX, CA, LA and the following countries: UK, Spain, Poland, Israel, Czech Republic, France, Belgium, Denmark, Italy, Cyprus, Austria, Greece, Slovenia, Australia, Germany, Estonia, New Zealand, Switzerland, Costa Rica, Guatemala, Canada, Malaysia, Poland, Singapore, Lithuania, Romania
Questions about this recall
Why was it recalled?
WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.
Which lots are affected?
1620005
Where was it sold?
WI, TX, CA, LA and the following countries: UK, Spain, Poland, Israel, Czech Republic, France, Belgium, Denmark, Italy, Cyprus, Austria, Greece, Slovenia, Australia, Germany, Estonia, New Zealand, Switzerland, Costa Rica, Guatemala, Canada, Malaysia, Poland, Singapore, Lithuania, Romania
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Waismed
All 3| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Nio-a NIO Intraosseous Device Adult recalled by Waismed | Waismed | Device Malfunction | Nationwide |
| Device | CLASS II | NIO Intraosseous Device Adult recalled by Waismed | Waismed | Device Malfunction | Nationwide |
Other Surgical Instruments recalls
All 3,891| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adaptable Beach Chair Beach Chair surgical patient positioner recalled | Quantum OPS | Device Malfunction | Nationwide |
| Device | CLASS II | Laparoscopic Prostatectomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Halyard Laparoscopy Comprehensive Pack II recalled | GRI Medical & Electronic Technology | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy/Laparotomy kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Gyn Laparoscopy Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |