CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report
Nitinol K-wire Blunt recalled by VB Spine
VB Spine is recalling Nitinol K-wire Blunt, distributed nationwide in the US. The recall was initiated on 2 Oct 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0440-2027
- FDA event ID
- 100141
- Recalling firm
- VB Spine, New York, NY
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 2 Oct 2026
- FDA classified
- 12 Nov 2026
- Reported
- 18 Nov 2026
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 2 Oct 2026 | Recall initiated by company | |
| 12 Nov 2026 | Classified by FDA | +41 days |
| 18 Nov 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for the K-wire to fracture during use resulting in procedural delay and possibility of a piece of fractured K-wire remaining in the patient's bone.
Product as listed by the FDA
NITINOL K-WIRE BLUNT; Catalog/Part Number: 48280232; used in spinal surgery
Where it was distributed
Domestic: Nationwide Distribution; International: Australia, Belgium, France, Ireland, Italy, Japan, Netherlands, Spain, United Kingdom;
Questions about this recall
Why was it recalled?
Potential for the K-wire to fracture during use resulting in procedural delay and possibility of a piece of fractured K-wire remaining in the patient's bone.
Which lots are affected?
Catalog/Part Number: 48280232; UDI-DI: 07613327096200; Serial/Lot Number(s) Affected: 25A459, 25A460, 25A461, 25A462, 25A463, 25A464, 25A465, 25A466, 25A467, 25A468, 25A469, 25A470, 25A471, 25A472, 25A473, 25A474, 25A475, 25A476, 25A477, 25A478, 25A479, 25A480, 25A481, 25A482, 25A483, 25A484, 25A485, 25A486, 25A487, 25A488, 25A489, 25A490, 25A491, 25A492, 25A493, 25A494, 25A495, 25A496, 25A497, 25A498, 25A499, 25A500, 25A501, 25A502, 25A503, 25A504, 25A505, 25A506, 25A507, 25A508, 25A509, 25A513, 25A512, 25A514, 25A515, 25A511, 25A510, 25A516, 25A517, 25A518, 25A519, 25A522, 25A525, 25A520, 25A521, 25A523, 25A524, 25A526;
Where was it sold?
Domestic: Nationwide Distribution; International: Australia, Belgium, France, Ireland, Italy, Japan, Netherlands, Spain, United Kingdom;
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Spinal Implants recalls
All 1,162| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline surgical convenience kits labeled as: Spinal CDS recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline surgical convenience kits labeled as recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |