CLASS II ONGOING Device recall · Reported 12 Feb 2025 · FDA Enforcement Report

Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 recalled

Nobel Biocare is recalling Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm, distributed nationwide in the US. The recall was initiated on 10 Jan 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesNobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213
Quantity recalled1,558 units (90 units - U.S.)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1101-2025
FDA event ID
96047
Recalling firm
Nobel Biocare, Goteborg
Classification
Class II
Status
Ongoing
Recall initiated
10 Jan 2025
FDA classified
5 Feb 2025
Reported
12 Feb 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility
Device type
Dental Devices

Timeline

10 Jan 2025Recall initiated by company
5 Feb 2025Classified by FDA+26 days
12 Feb 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

. Nobel Biocare N1 TiUltra TCC RP 4.0x9mm . Nobel Biocare N1 TiUltra TCC RP 4.0x11mm . Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Nobel Biocare N1 TiUltra TCC implants are intended for use as an endosseous dental implant in the maxilla or mandible for anchoring or supporting dental protheses to restore chewing function

Where it was distributed

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

Nationwide CaliforniaFloridaIllinoisMassachusettsNew JerseyNew YorkOhioPennsylvaniaUtahVirginiaWashington

Questions about this recall

Why was it recalled?

Due to manufacturing issue (unintended by-product of the injection molding process) a sharp pin that may puncture the blister packaging and breaching the sterile barrier of the finished product.

Which lots are affected?

Nobel Biocare N1 TiUltra TCC RP 4.0x 9mm Catalog Number: 300861 UDI-DI code: 07332747161731 Lot Numbers: 12253672 12253673 Nobel Biocare N1 TiUltra TCC RP 4.0x11mm Catalog Number: 300862 UDI-DI code: 07332747161748 Lot Numbers: 12253448 12254209 Nobel Biocare N1 TiUltra TCC RP 4.0x13mm Catalog Number: 300863 UDI-DI code: 07332747161755 Lot Numbers: 12253620 12254213

Where was it sold?

U.S. Nationwide distribution in the states of CA, FL, IL, MA, NJ, NY, OH, PA, UT, VA and WA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Nobel Biocare

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIN1 TiUltra TCC NP 3.5x13mm recalled by Nobel BiocareNobel BiocareOtherNationwide

Other Dental Devices recalls

All 426