CLASS II ONGOING Device recall · Reported 10 Sep 2025 · FDA Enforcement Report

NobelBiocare Tri-Lobe-compatible WP X Engaging Titanium Base recalled

Preat is recalling NobelBiocare Tri-Lobe-compatible WP X Engaging Titanium Base, distributed nationwide in the US. The recall was initiated on 25 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00842092169599. Lot: 277097
Quantity recalled11

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2514-2025
FDA event ID
97239
Recalling firm
Preat, Santa Maria, CA
Classification
Class II
Status
Ongoing
Recall initiated
25 Jun 2025
FDA classified
3 Sep 2025
Reported
10 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

25 Jun 2025Recall initiated by company
3 Sep 2025Classified by FDA+70 days
10 Sep 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

Product as listed by the FDA

NobelBiocare Tri-Lobe-compatible WP X 6.0mm Engaging Titanium Base, REF: 9000662

Where it was distributed

US Nationwide distribution in the states of NY, FL, PA, CA, OH.

Nationwide CaliforniaFloridaNew YorkOhioPennsylvania

Questions about this recall

Why was it recalled?

Engaging titanium base was manufactured 180 degrees around the post from where it should be located in relation to the connection geometry; if the flat is on the wrong side of the post, the crown will not fit onto the abutment in the correct orientation, resulting in restoration delay.

Which lots are affected?

UDI-DI: 00842092169599. Lot: 277097

Where was it sold?

US Nationwide distribution in the states of NY, FL, PA, CA, OH.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.