CLASS II ONGOING Device recall · Reported 16 Feb 2022 · FDA Enforcement Report
Novalung Sensor Box (F30000163) a component of Novalung recalled
Fresenius Medical Care Holdings is recalling Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated, distributed nationwide in the US. The recall was initiated on 7 Dec 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0562-2022
- FDA event ID
- 89369
- Recalling firm
- Fresenius Medical Care Holdings, Waltham, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Dec 2021
- FDA classified
- 4 Feb 2022
- Reported
- 16 Feb 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 7 Dec 2021 | Recall initiated by company | |
| 4 Feb 2022 | Classified by FDA | +59 days |
| 16 Feb 2022 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted communication between flow sensor and sensor box might lead to a disabled flow measurement and air bubble detection.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Novalung Sensor Box (F30000163) a component of Novalung Console (F30000162)- Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation and CO2 removal) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure, where other available treatment options have failed,
Where it was distributed
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ, NV, NY, OK, OR, PA, TN, TX, WV.
Nationwide AlabamaArizonaCaliforniaColoradoFloridaIllinoisMichiganNew JerseyNevadaNew YorkOklahomaOregonPennsylvaniaTennesseeTexasWest Virginia
Questions about this recall
Why was it recalled?
Error messages 206 (yellow) and 208 (red) technical failure, flow measurement during use of the Novalung System imply that the communication between flow sensor and sensor box is interrupted. An interrupted communication between flow sensor and sensor box might lead to a disabled flow measurement and air bubble detection.
Which lots are affected?
Serial Numbers: XCONUS0013 XCONUS0079 XCONUS0105 XCONUS0036 XCONUS0044 XCONUS0004 XCONUS0014 XCONUS0080 XCONUS0109 XCONUS0021 XCONUS0066 XCONUS0027 XCONUS0017 XCONUS0084 XCONUS0061 XCONUS0043 XCONUS0089 XCONUS0059 XCONUS0019 XCONUS0085 XCONUS0051 XCONUS0047 XCONUS0005 XCONUS0042 XCONUS0020 XCONUS0086 XCONUS0110 XCONUS0055 XCONUS0010 XCONUS0083 XCONUS0022 XCONUS0087 XCONUS0111 XCONUS0064 XCONUS0053 XCONUS0121 XCONUS0023 XCONUS0088 XCONUS0112 XCONUS0002 XCONUS0056 XCONUS0018 XCONUS0024 XCONUS0090 XCONUS0113 XCONUS0003 XCONUS0006 XCONUS0054 XCONUS0030 XCONUS0091 XCONUS0115 XCONUS0007 XCONUS0052 XCONUS0081 XCONUS0032 XCONUS0094 XCONUS0116 XCONUS0012 XCONUS0073 XCONUS0008 XCONUS0039 XCONUS0095 XCONUS0118 XCONUS0070 XCONUS0009 XCONUS0040 XCONUS0096 XCONUS0025 XCONUS0072 XCONUS0001 XCONUS0049 XCONUS0092 XCONUS0026 XCONUS0075 XCONUS0011 XCONUS0058 XCONUS0093 XCONUS0028 XCONUS0082 XCONUS0045 XCONUS0067 XCONUS0097 XCONUS0122 XCONUS0048 XCONUS0065 XCONUS0068 XCONUS0098 XCONUS0123 XCONUS0063 XCONUS0029 XCONUS0069 XCONUS0099 XCONUS0124 XCONUS0031 XCONUS0035 XCONUS0074 XCONUS0100 XCONUS0034 XCONUS0071 XCONUS0046 XCONUS0076 XCONUS0102 XCONUS0038 XCONUS0060 XCONUS0050 XCONUS0077 XCONUS0103 XCONUS0041 XCONUS0062 XCONUS0015 XCONUS0078 XCONUS0104 XCONUS0033 XCONUS0037 XCONUS0016 UDI: Prefix: 4057224 (+ serial number)
Where was it sold?
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, IL, MI, NJ, NV, NY, OK, OR, PA, TN, TX, WV.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Fresenius Medical Care Holdings
All 165| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | 5008X Blood Tubing Set recalled by Fresenius Medical Care | Fresenius Medical Care Holdings | Other | Nationwide |
| Device | CLASS II | Bicarby" Dialysate recalled by Fresenius Medical Care | Fresenius Medical Care Holdings | Packaging Defects | Nationwide |
| Device | CLASS II | Bicarby Dialysate recalled by Fresenius Medical Care | Fresenius Medical Care Holdings | Other | TX |
| Drug | CLASS II | Delflex, Dextrose Peritoneal Dialysis Solution recalled | Fresenius Medical Care Holdings | Sterility | Nationwide |
| Drug | CLASS II | Delflex, Dextrose Peritoneal Dialysis Solution recalled | Fresenius Medical Care Holdings | Sterility | Nationwide |
Other CPAP & Ventilators recalls
All 552| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS I | Astral 100/150 Resmed Astral 100/50 is a ventilator system recalled | ResMed | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |