CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report

OCS Liver Perfusion Set recalled by TransMedics

TransMedics is recalling OCS Liver Perfusion Set, distributed nationwide in the US. The recall was initiated on 15 May 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes1. Perfusion Set SN: 4470-4-24374; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24374/14D202711100*; 2. Perfusion Set SN: 4470-4-23845; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23845/14D202710235*; 3. Perfusion Set SN: 4470-4-24226; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24226/14D20271110$*; 4. Perfusion Set SN: 4470-4-24559; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24559/14D202711026*; 5. Perfusion Set SN: 4470-4-24636; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24636/14D202711022*; 6. Perfusion Set SN: 4470-4-24508; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24508/14D202711020*; 7. Perfusion Set SN: 4470-4-24384; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24384/14D202711022*; 8. Perfusion Set SN: 4470-4-23182; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23182/14D20271023%*; 9. Perfusion Set SN: 4470-4-23650; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23650/14D202710194*; 10. Perfusion Set SN: 4470-4-23593; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23593/14D20271019A*; 11. Perfusion Set SN: 4470-4-23471; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23471/14D202710195*; 12. Perfusion Set SN: 4470-4-22403; Description: OCS Liver Perfusion Set; EXP Date: 9/10/27; UDI: *+B00944705/$$+74470-4-22403/14D20270910+*; 13. Perfusion Set SN: 4470-4-22433; Description: OCS Liver Perfusion Set; EXP Date: 9/3/27; UDI: *+B00944705/$$+74470-4-22433/14D202709033*; 14. Perfusion Set SN: 4470-4-22303; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22303/14D20270820/*; 15. Perfusion Set SN: 4470-4-22280; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22280/14D202708201*;
Quantity recalled18

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0381-2027
FDA event ID
100063
Recalling firm
TransMedics, Andover, MA
Classification
Class II
Status
Ongoing
Recall initiated
15 May 2026
FDA classified
6 Nov 2026
Reported
18 Nov 2026
Reason category
Device Malfunction
Device type
Medical Software

Timeline

15 May 2026Recall initiated by company
6 Nov 2026Classified by FDA+175 days
18 Nov 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

OCS Liver Perfusion Set Model/Catalog Number: 3200 Software Version: rev 3 OCS Liver Perfusion Module included in the Liver Perfusion Set used for organ transport.

Where it was distributed

Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.

NationwideCaliforniaFloridaIllinoisMassachusettsMinnesotaNorth CarolinaNebraskaNew YorkTexas

Questions about this recall

Why was it recalled?

The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.

Which lots are affected?

1. Perfusion Set SN: 4470-4-24374; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24374/14D202711100*; 2. Perfusion Set SN: 4470-4-23845; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23845/14D202710235*; 3. Perfusion Set SN: 4470-4-24226; Description: OCS Liver Perfusion Set; EXP Date: 11/10/27; UDI: *+B00944705/$$+74470-4-24226/14D20271110$*; 4. Perfusion Set SN: 4470-4-24559; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24559/14D202711026*; 5. Perfusion Set SN: 4470-4-24636; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24636/14D202711022*; 6. Perfusion Set SN: 4470-4-24508; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24508/14D202711020*; 7. Perfusion Set SN: 4470-4-24384; Description: OCS Liver Perfusion Set; EXP Date: 11/2/27; UDI: *+B00944705/$$+74470-4-24384/14D202711022*; 8. Perfusion Set SN: 4470-4-23182; Description: OCS Liver Perfusion Set; EXP Date: 10/23/27; UDI: *+B00944705/$$+74470-4-23182/14D20271023%*; 9. Perfusion Set SN: 4470-4-23650; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23650/14D202710194*; 10. Perfusion Set SN: 4470-4-23593; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23593/14D20271019A*; 11. Perfusion Set SN: 4470-4-23471; Description: OCS Liver Perfusion Set; EXP Date: 10/19/27; UDI: *+B00944705/$$+74470-4-23471/14D202710195*; 12. Perfusion Set SN: 4470-4-22403; Description: OCS Liver Perfusion Set; EXP Date: 9/10/27; UDI: *+B00944705/$$+74470-4-22403/14D20270910+*; 13. Perfusion Set SN: 4470-4-22433; Description: OCS Liver Perfusion Set; EXP Date: 9/3/27; UDI: *+B00944705/$$+74470-4-22433/14D202709033*; 14. Perfusion Set SN: 4470-4-22303; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22303/14D20270820/*; 15. Perfusion Set SN: 4470-4-22280; Description: OCS Liver Perfusion Set; EXP Date: 8/20/27; UDI: *+B00944705/$$+74470-4-22280/14D202708201*;

Where was it sold?

Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 100063

More recalls from TransMedics

All 10

Other Medical Software recalls