CLASS II Device recall · Reported 4 Mar 2015 · FDA Enforcement Report
Offset Cup Impactor Universal recalled by Brainlab AG
Brainlab is recalling Offset Cup Impactor Universal, distributed nationwide in the US. The recall was initiated on 16 Jan 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1215-2015
- FDA event ID
- 70291
- Recalling firm
- Brainlab, Feldkirchen, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 16 Jan 2015
- FDA classified
- 26 Feb 2015
- Reported
- 4 Mar 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Hip & Knee Implants
Timeline
| 16 Jan 2015 | Recall initiated by company | |
| 26 Feb 2015 | Classified by FDA | +41 days |
| 4 Mar 2015 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)
Where it was distributed
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
Nationwide ArizonaCaliforniaIdahoIllinoisMarylandMichiganMinnesotaOhioOklahomaOregonTennesseeWashingtonWisconsinWest Virginia
Questions about this recall
Why was it recalled?
The recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Which lots are affected?
Model 52856 BL CURVED CUP INSERTER UNIVERSAL SHAFT; and MODEL 52856A BL OFFSET CUP IMPACTOR UNIVERSAL SHAFT, all serial/lot numbers distributed prior to January 13, 2015.
Where was it sold?
Worldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Brainlab
All 39| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Origin Data Management recalled by Brainlab AG | Brainlab | Device Malfunction | Nationwide |
| Device | CLASS II | Cirq Arm System recalled by Brainlab AG | Brainlab | Potency & Assay | TX |
| Device | CLASS II | ExacTrac Dynamic recalled by Brainlab AG | Brainlab | Device Malfunction | Nationwide |
| Device | CLASS II | ExacTrac Dynamic recalled by Brainlab AG | Brainlab | Device Malfunction | 20 states |
| Device | CLASS II | ExacTrac Dynamic recalled by Brainlab AG | Brainlab | Processing Deviation | Nationwide |
Other Hip & Knee Implants recalls
All 3,111| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Intellijoint Navigation System: Workstation G2 Configured recalled | Intellijoint Surgical | Device Malfunction | Nationwide |
| Device | CLASS II | Trinity Non-Surgical Isolation Gown, AAMI PB70 Level 2, X-Large recalled | New York Embroidery Studio | Packaging Defects | Nationwide |
| Device | CLASS II | Axle Pin Assy Large Elbow Cobalt Chrome recalled by Howmedica Osteonics | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Triathlon Hinge Insert Size 4 5612-p-411 Triathlon Hinge recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio Recessed Metal Backed Patella 3045 recalled | Howmedica Osteonics | Foreign Material | Nationwide |