CLASS II ONGOING Device recall · Reported 11 Jun 2025 · FDA Enforcement Report

Olympus Evis Exera Iii Duodenovideoscope recalled

Olympus Corporation of the Americas is recalling Olympus Evis Exera Iii Duodenovideoscope, distributed nationwide in the US. The recall was initiated on 5 May 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.
Quantity recalled12,195 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1911-2025
FDA event ID
96727
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
5 May 2025
FDA classified
5 Jun 2025
Reported
11 Jun 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

5 May 2025Recall initiated by company
5 Jun 2025Classified by FDA+31 days
11 Jun 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

Where it was distributed

Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.

Nationwide Georgia

Questions about this recall

Why was it recalled?

Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.

Which lots are affected?

UDI-DI: 04953170405563. Model Number: TJF-Q190V. All serial numbers.

Where was it sold?

Worldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, Georgia, Germany, Ghana, Guadeloupe, Guatemala, Honduras, Hong Kong, Iceland, India, Iraq, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Morocco, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Palestinian Authority, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Republic of the Congo, R¿union, Russia, Saudi Arabia, Senegal, Seychelles, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, Uruguay, Vietnam, Yemen.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Olympus Corporation of the Americas

All 423

Other Other Medical Devices recalls

All 13,908