CLASS II ONGOING Device recall · Reported 27 Nov 2024 · FDA Enforcement Report

Olympus GIF-1TH190 recalled by Olympus Corporation of the

Olympus Corporation of the Americas is recalling Olympus GIF-1TH190, distributed nationwide in the US. The recall was initiated on 18 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.
Quantity recalled1749 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0498-2025
FDA event ID
95600
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
18 Sep 2024
FDA classified
20 Nov 2024
Reported
27 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

18 Sep 2024Recall initiated by company
20 Nov 2024Classified by FDA+63 days
27 Nov 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/Catalog Number: GIF-1TH190 Software Version: N/A Product Description: This instrument is intended to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories (such as biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery.

Where it was distributed

Domestic: US Nationwide Distribution.

Nationwide

Questions about this recall

Why was it recalled?

To better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.

Which lots are affected?

Model: GIF-1TH190, UDI-DI: 04953170343360, Serial #: Less than 2446103.

Where was it sold?

Domestic: US Nationwide Distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Olympus Corporation of the Americas

All 423

Other Surgical Instruments recalls

All 3,891