CLASS II Device recall · Reported 21 Mar 2018 · FDA Enforcement Report
OneStep CPR Complete, Adult Multi Function Electrodes recalled
Zoll Medical is recalling OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 01 (, distributed nationwide in the US. The recall was initiated on 4 Aug 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0977-2018
- FDA event ID
- 79204
- Recalling firm
- Zoll Medical, Chelmsford, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 4 Aug 2017
- FDA classified
- 13 Mar 2018
- Reported
- 21 Mar 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Device Malfunction
- Device type
- Pacemakers & Defibrillators
Timeline
| 4 Aug 2017 | Recall initiated by company | |
| 13 Mar 2018 | Classified by FDA | +221 days |
| 21 Mar 2018 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult energy levels when used with the ZOLL R Series Defibrillator. Defective electrodes will display PEDIATRIC PADS IN USE on the R Series defibrillator. Using the affected product with the M Series Defibrillator poses no risk.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 01 (case of 8 single electrodes, UDI: 10847946016279) and 8900 0224 01 (single electrode, UDI: 00847946016272) Product Usage: Intended Use: Defibrillation, Cardioversion, Noninvasive Pacing, EGG Monitoring, CPR Sensor. For use with ZOLL¿ Defibrillators: R Series, M Series. By Trained Personnel only, Including: Physicians, Nurses, Paramedics, Emergency Medical Technicians, Cardiovascular Laboratory Technicians.
Where it was distributed
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and in the countries of Canada.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesota
Questions about this recall
Why was it recalled?
A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult energy levels when used with the ZOLL R Series Defibrillator. Defective electrodes will display PEDIATRIC PADS IN USE on the R Series defibrillator. Using the affected product with the M Series Defibrillator poses no risk.
Which lots are affected?
Lot Number: 2517A (Expiration date: 2019-06-24)
Where was it sold?
Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and in the countries of Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Zoll Medical
All 11| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | ZOLL Powerheart G5 AED recalled by ZOLL Medical | Zoll Medical | Temperature Abuse | Nationwide |
| Device | CLASS II | ZOLL Powerheart G5 AED recalled by ZOLL Medical | Zoll Medical | Temperature Abuse | Nationwide |
| Device | CLASS II | ZOLL Powerheart G5 AED recalled by ZOLL Medical | Zoll Medical | Temperature Abuse | Nationwide |
| Device | CLASS II | ZOLL Powerheart G5 AED recalled by ZOLL Medical | Zoll Medical | Temperature Abuse | Nationwide |
| Device | CLASS II | ZOLL Powerheart G5 AED, Semi-Automatic recalled by Zoll Medical | Zoll Medical | Device Malfunction | Nationwide |
Other Pacemakers & Defibrillators recalls
All 494| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DermaV Laser System recalled by Lutronic | Lutronic | Other | Nationwide |
| Device | CLASS II | Convenience kits containing select SKUs of Polycarbonate recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits: Loop Pack DYNJ61460C Open Heart recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Acl Pack DYNJ64161C Arthroscopy recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Bib AV Fistula Pack DYNJ58304B recalled | Medline Industries | Device Malfunction | Nationwide |