CLASS II ONGOING Device recall · Reported 14 Sep 2022 · FDA Enforcement Report
Optima Coil System recalled by Balt Usa
Balt USA is recalling Optima Coil System, distributed in Arizona, Florida. The recall was initiated on 26 Jul 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1680-2022
- FDA event ID
- 90724
- Recalling firm
- Balt USA, Irvine, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 26 Jul 2022
- FDA classified
- 6 Sep 2022
- Reported
- 14 Sep 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 26 Jul 2022 | Recall initiated by company | |
| 6 Sep 2022 | Classified by FDA | +42 days |
| 14 Sep 2022 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.
Product as listed by the FDA
Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
Where it was distributed
U.S.: AZ and FL O.U.S.: None
Questions about this recall
Why was it recalled?
Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.
Which lots are affected?
Model Number: OPTI0520CSF10 UDI-DI Code: 00818053026164 Lot Number: F220600470
Where was it sold?
U.S.: AZ and FL O.U.S.: None
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Balt USA
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | MEGA Ballast Distal Access Platform Mega Ballast recalled by Balt Usa | Balt USA | Device Malfunction | 12 states |
| Device | CLASS II | Optima Coil System REFs recalled by Balt Usa | Balt USA | Foreign Material | Nationwide |
| Device | CLASS II | Prestige Coil System REFs recalled by Balt Usa | Balt USA | Foreign Material | Nationwide |
| Device | CLASS II | PRES0413CPKPL Prestige Coil Prestige Plus Complex Pack recalled | Balt USA | Device Malfunction | 28 states |
| Device | CLASS II | Optima Coil System recalled by Balt Usa | Balt USA | Device Malfunction | Nationwide |
Other Heart & Vascular Devices recalls
All 6,220| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TufTex¿ Over-the-Wire Embolectomy Catheter recalled | LeMaitre Vascular | Other | Nationwide |
| Device | CLASS II | Endovascular Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Bard Ep Xt Decapolar Steerable, Diagnostic Ep Catheter, 6f recalled | Stryker Sustainability Solutions | Particulate Matter | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |