CLASS II ONGOING Device recall · Reported 2 Sep 2026 · FDA Enforcement Report

ORT300, operating room table recalled by IMRIS Imaging

Imris Imaging is recalling ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000, distributed nationwide in the US. The recall was initiated on 6 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376
Quantity recalled13 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3034-2026
FDA event ID
99596
Recalling firm
Imris Imaging, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
6 Aug 2026
FDA classified
26 Aug 2026
Reported
2 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Device Malfunction

Timeline

6 Aug 2026Recall initiated by company
26 Aug 2026Classified by FDA+20 days
2 Sep 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.

Nationwide

Questions about this recall

Why was it recalled?

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Which lots are affected?

114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99596

This product is part of a recall event; the main page for the event is ORT200, operating room table recalled by IMRIS Imaging.

ReportedClassRecallCompanyReasonWhere
CLASS IIORT200, operating room table recalled by IMRIS ImagingImris ImagingDevice MalfunctionNationwide

More recalls from Imris Imaging

All 3

Other Other Medical Devices recalls

All 13,908