CLASS II ONGOING Device recall · Reported 24 Dec 2025 · FDA Enforcement Report
Orthofix NorthStar OCT Navigation Surgical Technique Guide recalled
Seaspine Orthopedics is recalling Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation, distributed in 24 states. The recall was initiated on 17 Nov 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0937-2026
- FDA event ID
- 97973
- Recalling firm
- Seaspine Orthopedics, Carlsbad, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 17 Nov 2025
- FDA classified
- 17 Dec 2025
- Reported
- 24 Dec 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Surgical Instruments
Timeline
| 17 Nov 2025 | Recall initiated by company | |
| 17 Dec 2025 | Classified by FDA | +30 days |
| 24 Dec 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
Product as listed by the FDA
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355
Where it was distributed
US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
AlabamaArizonaCaliforniaColoradoConnecticutFloridaIllinoisIndianaMichiganMissouriMississippiNew JerseyNevadaNew YorkOklahomaOregonPennsylvaniaRhode IslandSouth CarolinaSouth Dakota
Questions about this recall
Why was it recalled?
The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.
Which lots are affected?
Surgical Technique Guide revisions: D0003926A, D0003926B, D0003926C, D0003926D. REF/UDI-DI/Lot: PC2-400012/10889981260033/MK81527A, MK81527ARR, MK81527ARRR, MK81574AR, MK81574AR1, MK93487A, MK93488ARR, MK93489A, MK93490A, MK93490ARR, MK93491AR, MK93491AR1, MK943488A, MK95882A, MK95882ARR, ML52083B, ML53136B, ML53188B, ML69777B, ML69778B, ML71257B, MM1001498D, MM1001499D, MM1001748D, MM1001749D, MM1001813D, MM1001902D, MM1001903D, MM1002025D, MM1002316D, MM1002328D, MM1002344D; PC2-400013/10889981260040/AL657979C, AL658549C, AL658549CRRR, AL658641C, AL661470C, AL671917D, AL683456F, AL702785F; PC2-400135/10889981260057/BK104899A, BK104899AR, BK104899ARR, BK104899ARRR, BK104899ARRR4, BK106822A, CR49020D, CR49023D, CR49060D, CR51550E, CR51972D, CR56472E; PC2-400140/10889981260064/BK104900A, BK104900AR, BK104900ARRR, BK106825A, BK107806A, BK107806ARRR, BK108366A, BK108366ARRR, CR49822D, CR50158D, CR51973D, CR57547E, CR59411E; PC2-400145/10889981260071/BK104901A, BK104901ARRR, BK106827A, CR49021C, CR49061C, CR51974C; PC2-400155/10889981260088/BK104902A, BK104902ARRR, BK106829A, CR49062C, CR51551C, CR51975C, CR57550E, CR59413E; PC2-400235/10889981260095/BK104903A, BK104903ARRR, BK106831A, BK106831ARRR, BK131159C, CR49620C, CR52371D, CR54729D, CR58013D, CR59414D; PC2-400240/10889981260101/BK104904A, BK104904ARRR, BK106833A, BK131160C, CR49621C, CR52372D, CR59415D; PC2-400245/10889981275495/no lots released; PC2-400345/10889981260118/BK104905A, BK104905ARRR, BK106834A, BK106834ARRR, BK131161C, CR49622C, CR52373D, CR54208D; PC2-400355/10889981260125/BK104906A, BK106835A, CR51764C, CR52853D
Where was it sold?
US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Seaspine Orthopedics
All 11| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | SeaSpine Polyaxial Head recalled by Seaspine Orthopedics | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Reef TA Inserter recalled by Seaspine Orthopedics | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Daytona Small Stature Ti Locking Cap, Small Stature, T25 recalled | Seaspine Orthopedics | Device Malfunction | Nationwide |
| Device | CLASS II | Daytona Small Stature CoCr LockingCap, Small Stature, T25, BT recalled | Seaspine Orthopedics | Device Malfunction | Nationwide |
Other Surgical Instruments recalls
All 3,921| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Sterile, synthetic, absorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | TruSystem 7500 surgical table recalled by Baxter Healthcare | Baxter Healthcare | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, natural, nonabsorbable recalled by Henry Schein | Henry Schein | Other | Nationwide |
| Device | CLASS II | Sterile, non-absorbable 7-0 blue polypropylene surgical recalled | Henry Schein | Other | Nationwide |