CLASS I Device recall · Reported 10 Sep 2014 · FDA Enforcement Report
Orthopedic Pack Basic recalled by Customed
Customed is recalling Orthopedic Pack Basic, distributed nationwide in the US. The recall was initiated on 20 May 2014 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2507-2014
- FDA event ID
- 68536
- Recalling firm
- Customed, Fajardo, PR
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 20 May 2014
- FDA classified
- 29 Aug 2014
- Reported
- 10 Sep 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Hip & Knee Implants
Timeline
| 20 May 2014 | Recall initiated by company | |
| 29 Aug 2014 | Classified by FDA | +101 days |
| 10 Sep 2014 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
ORTHOPEDIC PACK BASIC CODE 900-2830. CONTENTS: (3) TOWELS ABSORBENT 15" X 20" (1) COVER MAYO STAND REINFORCED (1) U-DRAPE 60" X 70" WITH TAPE SPLIT 6 X 21 SPLIT (1) DRAPE% ECONOMY 53" X 77" (4) DRAPE UTILITY WITH TAPE (1) BAG SUTURE FLORAL (1) SHEET SPLITW/ADHES 108" X 77" STD SMS (1) COVER TABLE REINFORCED 50" X 90" (1) STOCKINETTE IMPERVIOUS 14" X 48" (1) TIME OUT BEACON NON WOVEN (1) GOWN XL SMS IMPERVIOUS REINFORCED AAMI LEVEL Ill (1) POLISHER CAUTERY TIP (1) SKIN MARKER INK W/8 LABEL/TIME OUT/RULER (1) COUNTER NDLIBLADE 20C FOAM/MAG (1) PENCIL CAUTERY ROCKER SWITCH WITH HOLSTER (1) GOWN IMPERVIOUS EXTRA REINFORCED LGE T/WRAP AAMI LEVEL Il (1) BAG GLASSINE PLAIN (1) ELECTRODE DUAL DEPRES Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Where it was distributed
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Questions about this recall
Is Orthopedic Pack Basic recalled?
Yes. Customed recalled Orthopedic Pack Basic on 20 May 2014. The recall is Class I and its status is Terminated. Recall number Z-2507-2014.
Why was it recalled?
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Which lots are affected?
Code 900-2830, 3 lots 140111332 140211832 140412672
Where was it sold?
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Customed
All 682| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Catalog Number: 900277 Suction Catheter Tray 14FR With Sodium recalled | Customed | Sterility | PR |
| Device | CLASS II | Catalog Number: 9002940 Obstetrical Surgical Pack Used by medical recalled | Customed | Sterility | PR |
| Device | CLASS II | Catalog Number: 900010 Suture Removal KIT Used by medical staff to recalled | Customed | Sterility | PR |
| Device | CLASS II | Catalog Number: 9001933 Ophtalmic Surgical Pack Used recalled | Customed | Sterility | PR |
| Device | CLASS II | Catalog Number: 900169 Universal Surgical Pack Used by medical staff recalled | Customed | Sterility | PR |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Nitrosamine Impurity | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Lead & Heavy Metals | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |