CLASS II ONGOING Device recall · Reported 1 Oct 2025 · FDA Enforcement Report
Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight recalled
Ortoma AB is recalling Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight, distributed nationwide in the US. The recall was initiated on 28 Jul 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2615-2025
- FDA event ID
- 97416
- Recalling firm
- Ortoma AB, Goteborg, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 28 Jul 2025
- FDA classified
- 22 Sep 2025
- Reported
- 1 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Hip & Knee Implants
Timeline
| 28 Jul 2025 | Recall initiated by company | |
| 22 Sep 2025 | Classified by FDA | +56 days |
| 1 Oct 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Ortoma Inserter Adapter OTD: Inserter Adapter OTD Straight REF 30-201 Model Number: 2001 UDI-DI code: 07350137521019 Inserter Adapter OTD Pinnacle Straight REF 30-202 Model Number: 1001 UDI-DI code: 07350137520593 Reamer Adapter OTD REF 30-157 Model Number: 1003 UDI-DI code: 07350137520081 Inserter Adapter OTD REF 30-158 Model Number: 1004 UDI-DI code: 07350137520098 The Inserter Adapters are intended to be attached to a Cup Inserter Tool. The Cup Inserter Tool (impactor) is intended to be tracked by and navigated to orient and position the hip implant cup correctly relative to preoperative planned position and orientation.
Where it was distributed
Worldwide - U.S. Nationwide distribution in the state of FL and the countries of Japan, Sweden.
Questions about this recall
Why was it recalled?
Due to a weld breakage between the pin and inserter holder of the Inserter Adapter.
Which lots are affected?
Model Number: 2001 Catalog Number: 30-201 UDI-DI code: 07350137521019 Lot Numbers: 20244803 Model Number: 1001 Catalog Number: 30-202 UDI-DI code: 07350137520593 Lot Numbers: 20245103 20245104 20245109 Model Number: 1003 Catalog Number: 30-157 UDI-DI code: 07350137520081 Lot Numbers: 20242001 20242008 20245119 Model Number: 1004 Catalog Number: 30-158 UDI-DI code: 07350137520098 Only distributed O.U.S.
Where was it sold?
Worldwide - U.S. Nationwide distribution in the state of FL and the countries of Japan, Sweden.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Hip & Knee Implants recalls
All 3,123| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Unity Knee Asc Modular Handle included recalled by Corin Medical | Corin Medical | Device Malfunction | Nationwide |
| Device | CLASS II | Halyard Anterior Hip Pack recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard Anterior Hip Pack recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Hip Scope Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | TS Femoral Prep Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |