CLASS II ONGOING Device recall · Reported 27 Feb 2019 · FDA Enforcement Report
PageWriter TC50 Government Bundle, Product recalled
Philips North America is recalling PageWriter TC50 Government Bundle, Product 860429, distributed nationwide in the US. The recall was initiated on 9 Jan 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0860-2019
- FDA event ID
- 81957
- Recalling firm
- Philips North America, Andover, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 9 Jan 2019
- FDA classified
- 21 Feb 2019
- Reported
- 27 Feb 2019
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Surgical Instruments
Timeline
| 9 Jan 2019 | Recall initiated by company | |
| 21 Feb 2019 | Classified by FDA | +43 days |
| 27 Feb 2019 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
PageWriter TC50 Government Bundle, Product 860429, Software revisions up to and including A.07.05.22 to evaluate the electrocardiogram of adult and pediatric patients
Where it was distributed
Domestic distribution nationwide. Foreign distribution worldwide.
Questions about this recall
Is PageWriter TC50 Government Bundle, Product 860429 recalled?
Yes. Philips North America recalled PageWriter TC50 Government Bundle, Product 860429 on 9 Jan 2019. The recall is Class II and its status is Ongoing. Recall number Z-0860-2019.
Why was it recalled?
Philips PageWriter TC cardiograph batteries which have reached end of life may overheat, causing the case to melt and/or the device to ignite, which can cause injury to patient or nearby users.
Which lots are affected?
USN1800002 USN1800003 USN1800004
Where was it sold?
Domestic distribution nationwide. Foreign distribution worldwide.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips North America
All 531Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Sterility | Nationwide |